Two-beam pulsed Doppler cardiac output measurement: reproducibility and agreement with thermodilution.

D H Wong1, C K Mahutte.   

Abstract

Two observers used two-beam pulsed Doppler ultrasound, equipped with a suprasternal probe, to measure cardiac output (QtDopp) in 38 ICU patients who had pulmonary artery catheters and in 20 adult volunteers. The two-beam pulsed Doppler method enables one device to measure simultaneously both aortic blood velocity and aortic diameter. Each observer was blind to the other's measurement and to the thermodilution cardiac output measurement (Qttd). Linear regression of the mean of both observer's QtDopp on Qttd showed QtDopp = 0.90.Qttd + 0.01 (see = 1.54 L/min, r = .90). Bias (+/- SD), defined as mean (QtDopp - Qttd) difference, was -0.69 +/- 1.55 L/min. Interobserver agreement was more variable in patients than volunteers; mean (observer 1 - observer 2) difference was 0.14 +/- 1.30 L/min in ICU patients and -0.09 +/- 0.92 L/min in volunteers. Two-beam pulsed Doppler ultrasound is a simpler method of measuring QtDopp than previous pulsed Doppler methods which measure separately the aortic diameter by echocardiography. Although its agreement with Qttd is close to other Doppler methods and has acceptable interobserver reproducibility, its accuracy remains operator-dependent.

Entities:  

Mesh:

Year:  1990        PMID: 2180638     DOI: 10.1097/00003246-199004000-00017

Source DB:  PubMed          Journal:  Crit Care Med        ISSN: 0090-3493            Impact factor:   7.598


  3 in total

1.  Movement of pulmonary artery catheters.

Authors:  M S Chew; E Sloth; M Veien; K V Jensen; H H Kure; J M Hasenkam
Journal:  Heart Vessels       Date:  1996       Impact factor: 2.037

Review 2.  Cardiac output monitoring by echocardiography: should we pass on Swan-Ganz catheters?

Authors:  A C Perrino
Journal:  Yale J Biol Med       Date:  1993 Sep-Oct

3.  Cardiac output measurements via echocardiography versus thermodilution: A systematic review and meta-analysis.

Authors:  Yun Zhang; Yan Wang; Jing Shi; Zhiqiang Hua; Jinyu Xu
Journal:  PLoS One       Date:  2019-10-03       Impact factor: 3.240

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.