Scanning electron microscopic assessment of coating irregularities and their precursors in unexpanded durable polymer-based drug-eluting stents.

Mounir W Z Basalus1, Kenneth Tandjung, Thea van Westen, Hanim Sen, Pasquelle K N van der Jagt, Dirk W Grijpma, Aart A van Apeldoorn, Clemens von Birgelen.   

Abstract

OBJECTIVES: To assess and quantify coating irregularities on unexpanded and expanded durable polymer-based drug-eluting stents (DES) to gain insights into the origin of coating irregularities.
BACKGROUND: Previous scanning electron microscopy (SEM) studies in various expanded DES revealed differences in frequency and size of coating irregularities between DES types and specific distribution patterns, however, the origin of these irregularities is unclear.
METHODS: We assessed at bench side a total of 1,200 SEM images obtained in 30 DES samples (15 expanded and 15 unexpanded) of Cypher Select Plus, Taxus Liberté, Endeavor, Xience V, and resolute.
RESULTS: For most coating irregularities seen on expanded DES (72%; 23/32), a matching irregularity (n = 18/23) and/or its precursor (n = 11/23) was observed in unexpanded DES. Unexpanded Cypher select showed (small) crater lesions and cracks together with precursors of "peeling." On unexpanded Taxus Liberté, thinning of polymer, small bare metal areas, wrinkles, and one precursor type were found. Unexpanded endeavor showed cracks, small bare metal areas, crater lesions, and precursors of the latter. Unexpanded Xience V and resolute mainly revealed crater lesions and their precursors. On unexpanded versus expanded DES, there was no difference in measured frequency of coating irregularities and precursors (P = ns) with the exception of more bare metal areas on expanded Taxus Liberte (P = 0.01).
CONCLUSIONS: Most coating irregularities, or the potential to develop them, are inherent to the unexpanded DES. Important determinants of the formation of coating irregularities may be the stent geometry and the physical properties of the coating, while stent-balloon interaction plays no major role.
Copyright © 2011 Wiley Periodicals, Inc.

Entities:  

Mesh:

Substances:

Year:  2012        PMID: 21805599     DOI: 10.1002/ccd.23273

Source DB:  PubMed          Journal:  Catheter Cardiovasc Interv        ISSN: 1522-1946            Impact factor:   2.692


  5 in total

1.  Scanning electron microscopic analysis of polymer damage in drug-eluting stents during multiple stenting.

Authors:  Aiko Shimokado; Takashi Kubo; Hirotoshi Utsunomiya; Hisashi Yamasaki; Takashi Tanimoto; Yasushi Ino; Yong Liu; Ikuko Teraguchi; Yuichi Ozaki; Yasutsugu Shiono; Makoto Orii; Kunihiro Shimamura; Takashi Yamano; Tomoyuki Yamaguchi; Kumiko Hirata; Toshio Imanishi; Takashi Akasaka
Journal:  Int J Cardiovasc Imaging       Date:  2013-09-13       Impact factor: 2.357

2.  Clinical Impact of Stent Design.

Authors:  Rebecca L Noad; Colm G Hanratty; Simon J Walsh
Journal:  Interv Cardiol       Date:  2014-04

3.  Drug diffusion and biological responses of arteries using a drug-eluting stent with nonuniform coating.

Authors:  Noboru Saito; Yuhei Mori; Sayaka Uchiyama
Journal:  Med Devices (Auckl)       Date:  2016-03-17

4.  Development of a Bioactive Polymeric Drug Eluting Coronary Stent Coating Using Electrospraying.

Authors:  C M McKittrick; M J Cardona; R A Black; C McCormick
Journal:  Ann Biomed Eng       Date:  2019-08-22       Impact factor: 3.934

Review 5.  Cardiovascular stents: overview, evolution, and next generation.

Authors:  Setareh Borhani; Shadi Hassanajili; Seyed Hossein Ahmadi Tafti; Shahram Rabbani
Journal:  Prog Biomater       Date:  2018-09-10
  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.