Literature DB >> 21804316

In vitro and in vivo comparison of applanation tonometry and rebound tonometry in dogs.

Noriyuki Nagata1, Masashi Yuki, Takashi Hasegawa.   

Abstract

Intraocular pressure (IOP) evaluated by applanation tonometry via TONO-PEN XL (TP), and rebound tonometry via TonoVet (TV) were compared in enucleated canine eyes with varied pressure of the anterior chamber (AC) and in clinical cases. TV measured IOP values were lower than IOP measurements of TP in the enucleated eyes with 5-10 mmHg of AC (P<0.0001), though there was no significant difference in IOP values obtained with TP and TV on the pressure ranges of 15-20 mmHg. However, TP detected IOP values were lower than IOP measurements of TV in the eyes with over 25 mmHg of AC (P<0.0001). The results of clinical cases were similar to the enucleated eye model. There was no significant difference in IOP values obtained from TP and TV in dogs with normotensive eyes. IOP measurements of TP were lower than those of TV in glaucomatous eyes (P<0.0001). TV was a reliable tonometer for measurement of IOP in hypertensive eyes, whereas it was less accurate than TP in hypotensive eyes. The characteristics of TP and TV should be considered in the evaluation of IOP in practice.

Entities:  

Mesh:

Year:  2011        PMID: 21804316     DOI: 10.1292/jvms.11-0251

Source DB:  PubMed          Journal:  J Vet Med Sci        ISSN: 0916-7250            Impact factor:   1.267


  7 in total

1.  Validation of the Icare® TONOVET plus rebound tonometer in normal rabbit eyes.

Authors:  Shawna Gloe; Abby Rothering; Julie A Kiland; Gillian J McLellan
Journal:  Exp Eye Res       Date:  2019-06-12       Impact factor: 3.467

2.  Validation and comparison of four handheld tonometers in normal ex vivo canine eyes.

Authors:  Andrea L Minella; Julie A Kiland; Shawna Gloe; Gillian J McLellan
Journal:  Vet Ophthalmol       Date:  2020-06-01       Impact factor: 1.644

3.  Influence of Age on Ocular Biomechanical Properties in a Canine Glaucoma Model with ADAMTS10 Mutation.

Authors:  Joel R Palko; Hugh J Morris; Xueliang Pan; Christine D Harman; Kristin L Koehl; Kirk N Gelatt; Caryn E Plummer; András M Komáromy; Jun Liu
Journal:  PLoS One       Date:  2016-06-06       Impact factor: 3.240

4.  Comparative intraocular pressure measurements using three different rebound tonometers through in an ex vivo analysis and clinical trials in canine eyes.

Authors:  Jaeho Shim; Seonmi Kang; Yoonji Park; Sunhyo Kim; Seokmin Go; Eunji Lee; Kangmoon Seo
Journal:  Vet Ophthalmol       Date:  2020-06-05       Impact factor: 1.644

5.  Comparison among TonoVet, TonoVet Plus, Tono-Pen Avia Vet, and Kowa HA-2 portable tonometers for measuring intraocular pressure in dogs.

Authors:  João Victor Goulart Consoni Passareli; Felipe Franco Nascimento; Giovana José Garcia Estanho; Claudia Lizandra Ricci; Glaucia Prada Kanashiro; Rogério Giuffrida; Silvia Franco Andrade
Journal:  Vet World       Date:  2021-09-21

6.  Intraocular pressure measurements using the TONOVET® rebound tonometer: Influence of the probe-cornea distance.

Authors:  Blanche D Rodrigues; Fabiano Montiani-Ferreira; Mariza Bortolini; André T Somma; András M Komáromy; Peterson Triches Dornbusch
Journal:  Vet Ophthalmol       Date:  2020-10-17       Impact factor: 1.644

7.  The influence of the tonometer position on canine intraocular pressure measurements using the Tonovet® rebound tonometer.

Authors:  Juliana Kravetz de Oliveira; Fabiano Montiani-Ferreira; David L Williams
Journal:  Open Vet J       Date:  2018-03-09
  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.