Literature DB >> 21802153

Performance of two hearing screening protocols in NICU in Shanghai.

Zheng-Min Xu1, Wen-Xia Cheng, Xiao-Lin Yang.   

Abstract

OBJECTIVE: To study the sensitivity and specificity of targeted neonatal hearing screening for the single-session distortion product otoacoustic emissions (DPOAE) technique and the combined DPOAE/automated auditory brain-stem response (AABR) technique.
METHODS: 3000 high-risk newborns were studied at Children's Hospital of Fudan University. They were required to take two different screening procedures separately. The first procedure consisted of DPOAE alone and the second consisted of DPOAE combined with the AABR. Based upon the etiology in high-risk babies, they were divided into four groups. In group I there were 670 very-low-birth-weight (VLBW) newborns (1340 ears), and in group II there were 890 preterm babies (1780 ears). 850 babies (1700 ears) suffered from hyperbilirubinemia in group III, whereas 790 babies (1580 ears) with asphyxia were in group IV. The babies in groups II, III, and IV came from the neonatal intensive-care unit (NICU) of our hospital. The study protocols consisted of the DPOAE alone and DPOAE combined with AABR hearing screening at an age of less than 1 month, and a diagnostic stage at the age of 2 months.
RESULTS: With single-session DPOAE screening, the referral rate (8% of the NICU babies), the false-positive rate (4.96%) and the false-negative rate (0.8%) were higher. The different etiologies in NICU babies had significantly different referral rates (F-test, p<0.01). A 4.46% referral rate of hyperbilirubinemi babies was much lower. The combined DPOAE/AABR screening technique revealed a referral rate of 5.03%, a false-positive rate of 2% and a false-negative rate of 0.06%. The false-positive rate was well below the suggested 3% of the American Academy of Pediatric. Comparisons of the referral rate, false-positive rate and false-negative rate of two hearing screening protocols (DPOAE alone and combined DPOAE/AABR) revealed significant differences (t-test, p<0.05, p<0.01, p<0.01). 91 infants (3.03% of the NICU babies) who failed the combined DPOAE/AABR screening were confirmed on hearing impairment. Of 22 babies who passed DPOAE screening but failed the AABR screening had a severe to profound hearing loss based on classic ABR. These patients (24% of the NICU babies with hearing losses) with hyperbilirubinemia and asphyxia problems at newborn stage were diagnosed as auditory neuropathy based on evaluations of DPOAE screening passed, abnormal ABR and absent middle-ear muscle reflexes (MMR).
CONCLUSION: Our study demonstrates the use of a combination of DPOAE and AABR testing ensures high sensitivity and acceptable specificity, and predict the AN profile in NICU babies. Our efforts identified 22 NICU babies with auditory neuropathy who hopefully will benefit from early remediation of their hearing deficit.
Copyright © 2011 Elsevier Ireland Ltd. All rights reserved.

Entities:  

Mesh:

Year:  2011        PMID: 21802153     DOI: 10.1016/j.ijporl.2011.07.004

Source DB:  PubMed          Journal:  Int J Pediatr Otorhinolaryngol        ISSN: 0165-5876            Impact factor:   1.675


  6 in total

1.  The Hearing Status of Preterm Infant's ≤ 34 Weeks as Revealed by Otoacoustic Emissions (OAE) Screening and Diagnostic Brainstem Evoked Response Audiometry (BERA): A Tertiary Center Experience.

Authors:  Achal Gulati; Pirabu Sakthivel; Ishwar Singh; Siddarth Ramji
Journal:  Indian J Otolaryngol Head Neck Surg       Date:  2020-07-08

2.  The effects and outcomes of electrolyte disturbances and asphyxia on newborns hearing.

Authors:  Chun Liang; Qi Hong; Tao-Tao Jiang; Yan Gao; Xiao-Fang Yao; Xiao-Xing Luo; Xiu-Hui Zhuo; Jennifer B Shinn; Raleigh O Jones; Hong-Bo Zhao; Guang-Jin Lu
Journal:  Int J Pediatr Otorhinolaryngol       Date:  2013-05-04       Impact factor: 1.675

Review 3.  Preterm Hypoxic-Ischemic Encephalopathy.

Authors:  Krishna Revanna Gopagondanahalli; Jingang Li; Michael C Fahey; Rod W Hunt; Graham Jenkin; Suzanne L Miller; Atul Malhotra
Journal:  Front Pediatr       Date:  2016-10-20       Impact factor: 3.418

4.  Neonatal hearing screening in remote areas of China: a comparison between rural and urban populations.

Authors:  Wu Wenjin; Tang Xiangrong; Li Yun; Lü Jingrong; Chen Jianyong; Wang Xueling; Huang Zhiwu; Wu Hao
Journal:  J Int Med Res       Date:  2017-06-12       Impact factor: 1.671

5.  Neonatal Hearing Screening, with Otoacoustic Emission, among Normal Babies in a Northeastern Nigerian Hospital.

Authors:  Amina Abdullahi; Nurudeen Adebola Shofoluwe; Adamu Abubakar; Mala Bukar Sandabe; Hamman Ibrahim Garandawa; Mohammed Ibrahim Babatunde; Raheem O Quadri
Journal:  J West Afr Coll Surg       Date:  2022-06-22

6.  Outcomes of Automated Auditory Evoked Potential Performed in Different Settings and the Factors Associated with Referred Cases.

Authors:  Daniela Polo Camargo da Silva; Georgea Espíndola Ribeiro; Gustavo Leão Castilho; Jair Cortez Mantovani
Journal:  Int Arch Otorhinolaryngol       Date:  2017-10-25
  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.