Discrepancies between arterial oxygen saturation and functional oxygen saturation measured with pulse oximetry in very preterm infants.

Rhonda J Rosychuk1, Ann Hudson-Mason, Dianne Eklund, Thierry Lacaze-Masmonteil.   

Abstract

BACKGROUND: Discrepancies between pulse oximetry saturation (SpO(2)) and arterial saturation (SaO(2)) at low blood oxygenation values have been previously reported with significant variations among instruments and studies. Whether pulse oximeters that attenuate motion artifact are less prone to such discrepancies is not well known.
OBJECTIVE: To prospectively assess the agreement of the Masimo V4 pulse oximeter within the critical 85-95% SpO(2) target range. PATIENTS AND METHODS: For all consecutive babies with gestational age <33 weeks, postnatal age <7 days, and an umbilical arterial line, SpO(2) was measured continuously and SaO(2) analyzed on an as-needed basis. Bland-Altman techniques provided estimates of the difference (D = SaO(2) - SpO(2)), standard deviation (SD), and 95% limits of agreement (D ± 2*SD).
RESULTS: There were 1,032 measurements (114 babies) with SpO(2) between 85 and 95%. The 95% limits of agreement were -2.0 ± 5.8, -2.4 ± 9.2, and -1.9 ± 5.3 in the SpO(2) categories 85-95, 85-89, and 91-95%, respectively. For the SpO(2) categories 85-89% and 91-95%, only 52% (53/101) and 59% (523/886) of SpO(2) values, respectively, corresponded to the analogous SaO(2) categories. In the 85-89% SpO(2) category, SaO(2) was lower than 85% in 39 of the 101 (39%) measurements.
CONCLUSION: SaO(2) was lower on average than SpO(2) with an increased bias at lower saturation. The -2.4 ± 9.2 95% limits of agreement for SaO(2) - SpO(2) in the 85-89% SpO(2) category suggest that SpO(2) and SaO(2) are not interchangeable and intermittent SaO(2) assessments are warranted when the targeted SpO(2) is within this range.
Copyright © 2011 S. Karger AG, Basel.

Entities:  

Mesh:

Substances:

Year:  2011        PMID: 21791935     DOI: 10.1159/000326797

Source DB:  PubMed          Journal:  Neonatology        ISSN: 1661-7800            Impact factor:   4.035


  12 in total

Review 1.  Current concepts of oxygen therapy in neonates.

Authors:  Siddarth Ramji; Ola D Saugstad; Ashish Jain
Journal:  Indian J Pediatr       Date:  2014-10-18       Impact factor: 1.967

Review 2.  Intermittent Hypoxemia in Preterm Infants.

Authors:  Juliann M Di Fiore; Peter M MacFarlane; Richard J Martin
Journal:  Clin Perinatol       Date:  2019-06-15       Impact factor: 3.430

Review 3.  Intermittent hypoxemia and oxidative stress in preterm infants.

Authors:  Juliann M Di Fiore; Maximo Vento
Journal:  Respir Physiol Neurobiol       Date:  2019-05-14       Impact factor: 1.931

Review 4.  Hemoglobin oxygen saturation targets in the neonatal intensive care unit: Is there a light at the end of the tunnel? 1.

Authors:  Payam Vali; Mark Underwood; Satyan Lakshminrusimha
Journal:  Can J Physiol Pharmacol       Date:  2018-10-26       Impact factor: 2.273

Review 5.  Oxygen targeting in preterm infants: a physiological interpretation.

Authors:  S Lakshminrusimha; V Manja; B Mathew; G K Suresh
Journal:  J Perinatol       Date:  2014-10-30       Impact factor: 2.521

6.  Pulse Oximetry with Two Infrared Wavelengths without Calibration in Extracted Arterial Blood.

Authors:  Ohad Yossef Hay; Meir Cohen; Itamar Nitzan; Yair Kasirer; Sarit Shahroor-Karni; Yitzhak Yitzhaky; Shlomo Engelberg; Meir Nitzan
Journal:  Sensors (Basel)       Date:  2018-10-15       Impact factor: 3.576

7.  Critical Congenital Heart Disease Detection in the Screening Era: Do Not Neglect the Examination!

Authors:  Deepika Sankaran; Heather Siefkes; Frank F Ing; Satyan Lakshminrusimha; Francis R Poulain
Journal:  AJP Rep       Date:  2021-06-16

8.  Monitoring Oxygenation and Gas Exchange in Neonatal Intensive Care Units: Current Practice in the Netherlands.

Authors:  Ratna N G B Tan; Estelle E M Mulder; Enrico Lopriore; Arjan B Te Pas
Journal:  Front Pediatr       Date:  2015-11-03       Impact factor: 3.418

Review 9.  Pulse oximetry: fundamentals and technology update.

Authors:  Meir Nitzan; Ayal Romem; Robert Koppel
Journal:  Med Devices (Auckl)       Date:  2014-07-08

10.  Usual Care and Informed Consent in Clinical Trials of Oxygen Management in Extremely Premature Infants.

Authors:  Irene Cortés-Puch; Robert A Wesley; Michael A Carome; Robert L Danner; Sidney M Wolfe; Charles Natanson
Journal:  PLoS One       Date:  2016-05-18       Impact factor: 3.240

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.