Endurance and performance of two different concepts for left ventricular stimulation with bipolar epicardial leads in long-term follow-up.

H Burger1, J Kempfert, A van Linden, Z Szalay, M Schoenburg, T Walther, T Ziegelhoeffer.   

Abstract

BACKGROUND: Epicardial left ventricular (LV) leads represent an alternative for CRT therapy if transvenous lead implantation fails. Data on endurance, performance, the impact of the surgical approach (lateral minithoracotomy vs. median sternotomy simultaneously with other cardiac surgery), and the optimal technical concept (screw-in vs. suture-on) is limited.
METHODS: Over a period of 48 months we evaluated 130 consecutive patients with comparable characteristics. A total of 54 screw-in (MyoDex™ 1084T, SJM) and 76 suture-on (Capture Epi 4968, Medtronic) bipolar epicardial steroid-eluting LV leads were implanted either via a left lateral or a median thoracotomy. Sensing, pacing threshold, impedance and NYHA class were recorded at defined time points.
RESULTS: No surgery-related death or major complication was observed. At the time of implantation, the pacing threshold, sensing and NYHA class did not differ significantly between the two groups. The impedances of screw-in leads were significantly lower compared to those of suture-on leads. Suture-on leads showed a moderate initial drop in their pacing threshold but afterwards remained stable. Screw-in leads were characterized by a moderate but significant increase in the pacing threshold in the first year followed by a continuous decrease thereafter. Twenty-four months post-implantation no differences between both lead types could be detected. Sensing and NYHA class improved in both groups. The surgical approach had no significant impact on lead functionality.
CONCLUSION: Our study showed that the implantation of epicardial leads was safe with very low complication rates. There was no superior technical epicardial lead concept (screw-in vs. suture-on leads) and all epicardial leads demonstrated an excellent long-term performance and durability. Therefore, it seems that epicardial leads represent a good alternative to transvenous leads and surgeons should be encouraged to implant epicardial leads during concomitant cardiac surgery when the indications for CRT are present. Thieme Medical Publishers 333 Seventh Avenue, New York, NY 10001, USA.

Entities:  

Mesh:

Year:  2011        PMID: 21789760     DOI: 10.1055/s-0031-1280066

Source DB:  PubMed          Journal:  Thorac Cardiovasc Surg        ISSN: 0171-6425            Impact factor:   1.827


  4 in total

Review 1.  Recent advances in the optimization of cardiac resynchronization therapy.

Authors:  Satish Chandraprakasam; Gina G Mentzer
Journal:  Curr Heart Fail Rep       Date:  2015-02

Review 2.  Optimizing CRT - Do We Need More Leads and Delivery Methods.

Authors:  Pieter Martens; Frederik Hendrik Verbrugge; Wilfried Mullens
Journal:  J Atr Fibrillation       Date:  2015-04-30

3.  Implantation of a cardiac resynchronization therapy and defibrillation system using a robotic approach.

Authors:  Sana Amraoui; Louis Labrousse; Jean-Luc Jansens; Manav Sohal; Pierre Bordachar; Philippe Ritter
Journal:  HeartRhythm Case Rep       Date:  2015-07-13

4.  Epicardial left ventricular leads via minimally invasive technique: a role of steroid eluting leads.

Authors:  Etem Caliskan; Florian Fischer; Felix Schoenrath; Maximilian Y Emmert; Francesco Maisano; Volkmar Falk; Christoph T Starck; Tomas Holubec
Journal:  J Cardiothorac Surg       Date:  2017-11-08       Impact factor: 1.637

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.