Literature DB >> 21783110

Failure analysis of the fractured wires in sternal perichronal loops.

Jesús Chao1, Roberto Voces, Carmen Peña.   

Abstract

We report failure analysis of sternal wires in two cases in which a perichronal fixation technique was used to close the sternotomy. Various characteristics of the retrieved wires were compared to those of unused wires of the same grade and same manufacturer and with surgical wire specifications. In both cases, wire fracture was un-branched and transgranular and proceeded by a high cycle fatigue process, apparently in the absence of corrosion. However, stress anlysis indicates that the effective stress produced during strong coughing is lower than the yield strength. Our findings suggest that in order to reduce the risk for sternal dehiscence, the diameter of the wire used should be increased.
Copyright © 2011 Elsevier Ltd. All rights reserved.

Entities:  

Mesh:

Year:  2011        PMID: 21783110     DOI: 10.1016/j.jmbbm.2011.02.015

Source DB:  PubMed          Journal:  J Mech Behav Biomed Mater        ISSN: 1878-0180


  3 in total

1.  Fatal bleeding due to sternal steel wire fracture following open-heart surgery.

Authors:  Wenhui Gong; Xiaofeng Ye; Zhe Wang; Sen Li; Qiang Zhao
Journal:  J Thorac Dis       Date:  2015-10       Impact factor: 2.895

2.  Wire fracture in postoperative Nuss procedure: a problem that cannot be ignored.

Authors:  Yunxuan Zhang; Qiang Chen; Yarui Luo; Chenyu Sun; Meng Chen; Na Wu; Yimin Xie
Journal:  Transl Pediatr       Date:  2021-03

3.  Comparison of polyetheretherketone cables and stainless steel wires for sternal fixation after median sternotomy.

Authors:  Duanchao Shu; Jianpeng Li; Yixin Zhao; Yan Yang
Journal:  J Int Med Res       Date:  2021-09       Impact factor: 1.671

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.