Literature DB >> 21778844

Comparison of removal torques for laser-treated titanium implants with anodized implants.

Sang-Hwan Kang1, Sung-Am Cho.   

Abstract

PURPOSE: It is difficult to draw any firm conclusions regarding possible differences in implant surface performance in a clinical situation. The purpose of this study was to compare the removal torques of laser-treated (LT) surface of dental implants with TiUnite, which has porous rough anodized surface, evaluating their ability to enhance the bond strength between a titanium substrate and rabbit femur.
METHOD: Two types of dental implants, LT (test, CP titanium grade 4) and commercial porous TiUnite (control, CP titanium grade 4) implants (length, 7 mm; diameter, 3.75 mm) were implanted on the right and left distal femoral metaphyses of 5 adult rabbits weighing 3.0 to 4.0 kg. After 8 weeks, removal torque was measured using a digital torque gauge.
RESULTS: The mean (SD) removal torque was 32.83 (6.15) and 48.59 (8.07) N · cm for the control (anodized) and test (LT) screws, respectively. P = 0.0133 by paired t-test.
CONCLUSIONS: The removal torque of the LT titanium implant embedded in rabbit femoral metaphysis for 8 weeks was stronger than that of the anodized titanium implant.

Entities:  

Mesh:

Substances:

Year:  2011        PMID: 21778844     DOI: 10.1097/SCS.0b013e31821d4d98

Source DB:  PubMed          Journal:  J Craniofac Surg        ISSN: 1049-2275            Impact factor:   1.046


  5 in total

1.  Laser-treated stainless steel mini-screw implants: 3D surface roughness, bone-implant contact, and fracture resistance analysis.

Authors:  He-Kyong Kang; Tien-Min Chu; Paul Dechow; Kelton Stewart; Hee-Moon Kyung; Sean Shih-Yao Liu
Journal:  Eur J Orthod       Date:  2015-04-23       Impact factor: 3.075

2.  Comparison of early osseointegration between laser-treated/acid-etched and sandblasted/acid-etched titanium implant surfaces.

Authors:  Mingdeng Rong; Haibin Lu; Lei Wan; Xueyang Zhang; Xi Lin; Shaobing Li; Lei Zhou; Yingtao Lv; Yuan Su
Journal:  J Mater Sci Mater Med       Date:  2018-03-30       Impact factor: 3.896

3.  Comparison of removal torques between laser-etched and modified sandblasted acid-etched Ti implant surfaces in rabbit tibias.

Authors:  Kyung-Soon Park; Abdel Ghani Ibrahim Al Awamleh; Sung-Am Cho
Journal:  J Adv Prosthodont       Date:  2018-02-12       Impact factor: 1.904

4.  Biomechanical and histological evaluation of four different titanium implant surface modifications: an experimental study in the rabbit tibia.

Authors:  José Luis Calvo-Guirado; Marta Satorres; Bruno Negri; Piedad Ramirez-Fernandez; Jose Eduardo Maté-Sánchez de Val; Jose Eduardo Maté-Sánchez; Rafael Delgado-Ruiz; Gerardo Gomez-Moreno; Marcus Abboud; Georgios E Romanos
Journal:  Clin Oral Investig       Date:  2013-10-18       Impact factor: 3.573

5.  Comparison of removal torques between laser-treated and SLA-treated implant surfaces in rabbit tibiae.

Authors:  Nam-Seok Kang; Lin-Jie Li; Sung-Am Cho
Journal:  J Adv Prosthodont       Date:  2014-08-14       Impact factor: 1.904

  5 in total