Monitoring of noninvasive ventilation by built-in software of home bilevel ventilators: a bench study.

Olivier Contal1, Laurence Vignaux2, Christophe Combescure3, Jean-Louis Pepin4, Philippe Jolliet5, Jean-Paul Janssens6.   

Abstract

BACKGROUND: Current bilevel positive-pressure ventilators for home noninvasive ventilation (NIV) provide physicians with software that records items important for patient monitoring, such as compliance, tidal volume (Vt), and leaks. However, to our knowledge, the validity of this information has not yet been independently assessed.
METHODS: Testing was done for seven home ventilators on a bench model adapted to simulate NIV and generate unintentional leaks (ie, other than of the mask exhalation valve). Five levels of leaks were simulated using a computer-driven solenoid valve (0-60 L/min) at different levels of inspiratory pressure (15 and 25 cm H(2)O) and at a fixed expiratory pressure (5 cm H(2)O), for a total of 10 conditions. Bench data were compared with results retrieved from ventilator software for leaks and Vt.
RESULTS: For assessing leaks, three of the devices tested were highly reliable, with a small bias (0.3-0.9 L/min), narrow limits of agreement (LA), and high correlations (R(2), 0.993-0.997) when comparing ventilator software and bench results; conversely, for four ventilators, bias ranged from -6.0 L/min to -25.9 L/min, exceeding -10 L/min for two devices, with wide LA and lower correlations (R(2), 0.70-0.98). Bias for leaks increased markedly with the importance of leaks in three devices. Vt was underestimated by all devices, and bias (range, 66-236 mL) increased with higher insufflation pressures. Only two devices had a bias < 100 mL, with all testing conditions considered.
CONCLUSIONS: Physicians monitoring patients who use home ventilation must be aware of differences in the estimation of leaks and Vt by ventilator software. Also, leaks are reported in different ways according to the device used.

Entities:  

Mesh:

Year:  2011        PMID: 21778253     DOI: 10.1378/chest.11-0485

Source DB:  PubMed          Journal:  Chest        ISSN: 0012-3692            Impact factor:   9.410


  20 in total

Review 1.  Basic Concepts for Tidal Volume and Leakage Estimation in Non-Invasive Ventilation.

Authors:  Manel Luján; Cristina Lalmolda; Begüm Ergan
Journal:  Turk Thorac J       Date:  2019-04-01

2.  Nasal versus oronasal continuous positive airway pressure masks for obstructive sleep apnea: is this really a key point of effectiveness?

Authors:  Antonio M Esquinas; Zeynep Zeren Ucar; Cenk Kirakli
Journal:  Sleep Breath       Date:  2013-04-08       Impact factor: 2.816

3.  Nasal versus oronasal continuous positive airway pressure masks for obstructive sleep apnea: is this really a key point of effectiveness?

Authors:  Jessie P Bakker; Alister M Neill; Angela J Campbell
Journal:  Sleep Breath       Date:  2013-04-10       Impact factor: 2.816

4.  And the Doctor Answers: "Dream, Dream, I Will Be the Guardian of Your Breathing…".

Authors:  Claudio Rabec; Jésus Gonzalez-Bermejo; Christophe Perrin; Bruno Langevin; Jean-Louis Pepin; Daniel Rodenstein; Jean Paul Janssens
Journal:  J Clin Sleep Med       Date:  2016-08-15       Impact factor: 4.062

5.  Noninvasive auto-titrating ventilation (AVAPS-AE) versus average volume-assured pressure support (AVAPS) ventilation in hypercapnic respiratory failure patients.

Authors:  Gul Gursel; Avsar Zerman; Burcu Basarik; Kamil Gonderen; Muge Aydogdu; Serriyye Memmedova
Journal:  Intern Emerg Med       Date:  2018-03-06       Impact factor: 3.397

6.  Polygraphic respiratory events during sleep with noninvasive ventilation in children: description, prevalence, and clinical consequences.

Authors:  Valeria Caldarelli; Jean Christian Borel; Sonia Khirani; Adriana Ramirez; Renato Cutrera; Jean-Louis Pépin; Brigitte Fauroux
Journal:  Intensive Care Med       Date:  2013-01-24       Impact factor: 17.440

7.  Performance characteristics of seven bilevel mechanical ventilators in pressure-support mode with different cycling criteria: a comparative bench study.

Authors:  Yuqing Chen; Kewen Cheng; Xin Zhou
Journal:  Med Sci Monit       Date:  2015-01-26

8.  Combined effects of leaks, respiratory system properties and upper airway patency on the performance of home ventilators: a bench study.

Authors:  Kaixian Zhu; Claudio Rabec; Jésus Gonzalez-Bermejo; Sébastien Hardy; Sami Aouf; Pierre Escourrou; Gabriel Roisman
Journal:  BMC Pulm Med       Date:  2017-11-21       Impact factor: 3.317

9.  Sleep increases leaks and asynchronies during home noninvasive ventilation: a polysomnographic study.

Authors:  Sergi Martí; Alex Ferré; Gabriel Sampol; Mercedes Pallero; Odile Romero; Jaume Ferrer; Júlia Sampol
Journal:  J Clin Sleep Med       Date:  2022-01-01       Impact factor: 4.062

10.  Effect of dynamic random leaks on the monitoring accuracy of home mechanical ventilators: a bench study.

Authors:  Ana Sogo; Jaume Montanyà; Eduard Monsó; Lluís Blanch; Xavier Pomares; Manel Lujàn
Journal:  BMC Pulm Med       Date:  2013-12-10       Impact factor: 3.317

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.