Jorge Calvo1, María Eliecer Cano, Cristina Pitart, Francesc Marco, José Manuel Rodríguez-Martínez, Alvaro Pascual, Luis Martínez-Martínez.
Abstract
The accuracy of the MicroScan WalkAway, BD Phoenix, and Vitek-2 systems for susceptibility testing of quinolones and aminoglycosides against 68 enterobacteria containing qnrB, qnrS, and/or aac(6 ')-Ib-cr was evaluated using reference microdilution. Overall, one very major error (0.09%), 6 major errors (0.52%), and 45 minor errors (3.89%) were noted.Entities:
Mesh:
Substances:
Year: 2011 PMID: 21775549 PMCID: PMC3165611 DOI: 10.1128/JCM.00563-11