Literature DB >> 21772396

Refractive index measured by laser beam displacement at λ=1064 nm for solvents and deuterated solvents.

David P Shelton1.   

Abstract

The refractive index of a liquid is determined with 0.0003 accuracy from measurements of laser beam displacement by a liquid-filled standard 10 mm spectrophotometer cell. The apparatus and methods are described and the results of measurements at λ=1064 nm and T=25.0 °C for 30 solvents and deuterated solvents are presented. Several sources of potential systematic errors as large as 0.003 are identified, the most important being the curvature of the liquid cell windows. The measurements are analyzed accounting for the significant imperfections of the apparatus.
© 2011 Optical Society of America

Year:  2011        PMID: 21772396     DOI: 10.1364/AO.50.004091

Source DB:  PubMed          Journal:  Appl Opt        ISSN: 1559-128X            Impact factor:   1.980


  2 in total

1.  Optofluidic Fabry-Pérot Micro-Cavities Comprising Curved Surfaces for Homogeneous Liquid Refractometry-Design, Simulation, and Experimental Performance Assessment.

Authors:  Noha Gaber; Yasser M Sabry; Frédéric Marty; Tarik Bourouina
Journal:  Micromachines (Basel)       Date:  2016-04-07       Impact factor: 2.891

2.  Evaluation of femtosecond laser-assisted anterior capsulotomy in the presence of ophthalmic viscoelastic devices (OVDs).

Authors:  Hassan Mansoor; Yu-Chi Liu; Yoke Rung Wong; Nyein C Lwin; Xin Y Seah; Jodhbir S Mehta
Journal:  Sci Rep       Date:  2020-12-09       Impact factor: 4.379

  2 in total