A hand-held instrument to maintain steady tissue contact during probe-based confocal laser endomicroscopy.

Win Tun Latt1, Richard C Newton, Marco Visentini-Scarzanella, Christopher J Payne, David P Noonan, Jianzhong Shang, Guang-Zhong Yang.   

Abstract

Probe-based confocal laser endomicroscopy (pCLE) provides high-resolution in vivo imaging for intraoperative tissue characterization. Maintaining a desired contact force between target tissue and the pCLE probe is important for image consistency, allowing large area surveillance to be performed. A hand-held instrument that can provide a predetermined contact force to obtain consistent images has been developed. The main components of the instrument include a linear voice coil actuator, a donut load-cell, and a pCLE probe. In this paper, detailed mechanical design of the instrument is presented and system level modeling of closed-loop force control of the actuator is provided. The performance of the instrument has been evaluated in bench tests as well as in hand-held experiments. Results demonstrate that the instrument ensures a consistent predetermined contact force between pCLE probe tip and tissue. Furthermore, it compensates for both simulated physiological movement of the tissue and involuntary movements of the operator's hand. Using pCLE video feature tracking of large colonic crypts within the mucosal surface, the steadiness of the tissue images obtained using the instrument force control is demonstrated by confirming minimal crypt translation.

Entities:  

Mesh:

Year:  2011        PMID: 21768038     DOI: 10.1109/TBME.2011.2162064

Source DB:  PubMed          Journal:  IEEE Trans Biomed Eng        ISSN: 0018-9294            Impact factor:   4.538


  11 in total

1.  Robot-assisted transvaginal peritoneoscopy using confocal endomicroscopy: a feasibility study in a porcine model.

Authors:  Richard C Newton; David P Noonan; Valentina Vitiello; James Clark; Christopher J Payne; Jianzhong Shang; Mikael Sodergren; Ara Darzi; Guang-Zhong Yang
Journal:  Surg Endosc       Date:  2012-04-26       Impact factor: 4.584

2.  Force adaptive robotically assisted endomicroscopy for intraoperative tumour identification.

Authors:  Petros Giataganas; Michael Hughes; Guang-Zhong Yang
Journal:  Int J Comput Assist Radiol Surg       Date:  2015-04-23       Impact factor: 2.924

3.  A Novel Semi-Autonomous Control Framework for Retina Confocal Endomicroscopy Scanning.

Authors:  Zhaoshuo Li; Mahya Shahbazi; Niravkumar Patel; Eimear O' Sullivan; Haojie Zhang; Khushi Vyas; Preetham Chalasani; Peter L Gehlbach; Iulian Iordachita; Guang-Zhong Yang; Russell H Taylor
Journal:  Rep U S       Date:  2020-01-27

4.  Hybrid position/force control of an active handheld micromanipulator for membrane peeling.

Authors:  Trent S Wells; Sungwook Yang; Robert A MacLachlan; Louis A Lobes; Joseph N Martel; Cameron N Riviere
Journal:  Int J Med Robot       Date:  2015-05-11       Impact factor: 2.547

5.  An Ungrounded Hand-Held Surgical Device Incorporating Active Constraints with Force-Feedback.

Authors:  Christopher J Payne; Ka-Wai Kwok; Guang-Zhong Yang
Journal:  Rep U S       Date:  2013-11-07

6.  A Hand-held Instrument for in vivo Probe-based Confocal Laser Endomicroscopy during Minimally Invasive Surgery.

Authors:  Win Tun Latt; Tou Pin Chang; Aimee Di Marco; Philip Pratt; Ka-Wai Kwok; James Clark; Guang-Zhong Yang
Journal:  Rep U S       Date:  2012-10-12

7.  Hybrid Robot-assisted Frameworks for Endomicroscopy Scanning in Retinal Surgeries.

Authors:  Zhaoshuo Li; Mahya Shahbazi; Niravkumar Patel; Eimear O' Sullivan; Haojie Zhang; Khushi Vyas; Preetham Chalasani; Anton Deguet; Peter L Gehlbach; Iulian Iordachita; Guang-Zhong Yang; Russell H Taylor
Journal:  IEEE Trans Med Robot Bionics       Date:  2020-04-16

8.  Toward Hybrid Position/Force Control for an Active Handheld Micromanipulator.

Authors:  Trent S Wells; Robert A MacLachlan; Cameron N Riviere
Journal:  IEEE Int Conf Robot Autom       Date:  2014-05

9.  A cable-driven parallel manipulator with force sensing capabilities for high-accuracy tissue endomicroscopy.

Authors:  Kiyoteru Miyashita; Timo Oude Vrielink; George Mylonas
Journal:  Int J Comput Assist Radiol Surg       Date:  2018-03-07       Impact factor: 2.924

10.  Use of endoscopic distal attachment cap to enhance image stabilization in probe-based confocal laser endomicroscopy in colorectal lesions.

Authors:  Vivian Ussui; Can Xu; Julia E Crook; Nancy N Diehl; Joy Hardee; Estela G Staggs; Muhammad W Shahid; Michael B Wallace
Journal:  Endosc Int Open       Date:  2015-07-21
View more