[Accuracy of a continuous glucose monitoring system in detection of blood glucose during oral glucose tolerance test].

Zhi Chen1, Jie Shen, Ling-ling Xu, Xia-jun Fu, Ji-min Li, Yuan-yuan Ma.   

Abstract

OBJECTIVE: To evaluate the accuracy of continuous glucose monitoring system (CGMS) during oral glucose tolerance test (OGTT) in the detection of blood glucose changes in glucose stress condition.
METHODS: Forty-nine out-patients with fasting plasma glucose of 3.9-11.0 mmol/L underwent continuous blood glucose monitoring using CGMS for 3 days, and OGTT was conducted on the third day. The venous blood glucose was measured at 0, 30, 60, 90, and 120 min after oral glucose intake, and the accuracy of CGMS during OGTT was evaluated.
RESULTS: The correlation indices between CGMS values and the venous blood glucose values during the entire OGTT and in phases of stable, rapidly rising and falling glucose levels were 0.928, 0.901, 0.924 and 0.902, respectively (P<0.001). Clarke error-grid analysis showed that more than 95% of the measured results fell into the A and B zones.
CONCLUSION: CGMS values show good consistency with venous blood glucose values measured during OGTT. CGMS is accurate in detection of rapidly changing blood glucose during OGTT.

Entities:  

Mesh:

Substances:

Year:  2011        PMID: 21764709

Source DB:  PubMed          Journal:  Nan Fang Yi Ke Da Xue Xue Bao        ISSN: 1673-4254


  2 in total

Review 1.  Continuous glucose monitoring system and new era of early diagnosis of diabetes in high risk groups.

Authors:  Ashraf Soliman; Vincenzo DeSanctis; Mohamed Yassin; Rania Elalaily; Nagwa E Eldarsy
Journal:  Indian J Endocrinol Metab       Date:  2014-05

2.  Glucose time series complexity as a predictor of type 2 diabetes.

Authors:  Carmen Rodríguez de Castro; Luis Vigil; Borja Vargas; Emilio García Delgado; Rafael García Carretero; Julián Ruiz-Galiana; Manuel Varela
Journal:  Diabetes Metab Res Rev       Date:  2016-06-30       Impact factor: 4.876

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.