A practical, single-view alternative to Stenver's for plain radiographic unilateral and bilateral post-cochlear implant position check.

Robert Harris1, Chris Pepper, Lee Dennis, Phillip Rich, David Selvadurai.   

Abstract

INTRODUCTION: 'Stenver's' is the standard plain radiograph view to check the electrode position after cochlear implantation. However, a reproducible alignment for intra-individual comparison of electrode position using a true Stenver's alignment is not always straightforward to achieve, particularly for inexperienced radiographers, or on non-compliant children. In addition, two ionizing exposures in two different positions are required for bilateral cochlear implants. AIMS: To assess the efficacy of other, more simple, single-exposure radiographs for the assessment of electrode position.
METHODS: Dummy electrodes and receiver units were bilaterally implanted into a cadaver. Plain radiographs in Stenver's views, and plain antero-posterior (AP) midline radiographic single views incorporating both implants were obtained with the tube at a range of angles tilted caudal and cranial to the perpendicular. Five internationally renowned cochlear implant surgeons were each asked if each radiograph gave adequate information for unilateral and for bilateral implants and they were asked to list in order of their favoured top three views.
RESULTS: No surgeon thought that a single-exposure Stenver's view was adequate for assessment of the contra-lateral side. Consensus was that all AP views were fit for the purpose, with no preference given between each of the AP views. The ipsi-lateral Stenver's was considered to give better depth of insertion information.
CONCLUSION: There is no apparent advantage of caudal and cranial tilt angles over a straight perpendicular AP. A single-view AP radiograph is an alternative to Stenver's view for a post-unilateral and post-simultaneous bilateral cochlear implant check.

Entities:  

Mesh:

Year:  2011        PMID: 21756460     DOI: 10.1179/146701010X486408

Source DB:  PubMed          Journal:  Cochlear Implants Int        ISSN: 1467-0100


  1 in total

1.  Measurement of Cochlear Implant Electrode Position From Intraoperative Post-insertion Skull Radiographs: A Validation Study.

Authors:  Maja Svrakic; David R Friedmann; Phillip M Berman; Adam J Davis; J Thomas Roland; Mario A Svirsky
Journal:  Otol Neurotol       Date:  2015-09       Impact factor: 2.311

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.