Yimin Wang, Amani A Fawzi, Ou Tan, Xinbo Zhang, David Huang.
Abstract
We used Doppler Fourier-domain optical coherence tomography (OCT) to investigate flicker-induced changes of total retinal blood flow. Total retinal blood flow was measured by summing flows in veins imaged in double-circular scans around the optic disc. In 3 healthy volunteers, total retinal blood flow was measured before and 10-15 seconds after 30 seconds of flicker stimulation. The average blood flow increased 22.2% (p = 0.002). The total venous and arterial vessel cross-sectional area increased 11.3% (p < 0.001) and +2.7% (p = 0.28) respectively. The average venous and arterial flow velocity were calculated indirectly by dividing total retinal blood flow by total venous and arterial cross-sectional areas. They also increased by 8.8% (p = 0.046) and 18.3% (p = 0.004), respectively. These results show that human retinal blood flow increases after visible flicker stimulation, and this could be measured with OCT.Entities:
Keywords: (170.3880) Medical and biological imaging; (170.4470) Ophthalmology; (170.4500) Optical coherence tomography; (280.2490) Flow diagnostics
Year: 2011 PMID: 21750763 PMCID: PMC3130572 DOI: 10.1364/BOE.001852
Source DB: PubMed Journal: Biomed Opt Express ISSN: 2156-7085 Impact factor: 3.732
Fig. 1Experiment setup of light stimulation with ring LED array. EU, Electrical unit; OL, Objective lens
Fig. 2Blood volume flow rate versus blood vessel diameter. Results are on a log-log scale. Solid line: best-fit result of linear regression (P < 0.001; R2 = 0.88). Dotted lines: 95% confidence interval limits to the fitted line (solid).
Fig. 3Variation of total retinal blood flow at different stimulus brightness. The vertical line on each point shows the standard deviation (SD) of repeat measurements at each setting.
Change in total retinal blood flow with flicker stimulation
| Subject | Total Retinal Blood Flow | ||
|---|---|---|---|
| Base line (µl/min) (Mean ± SD) | Stimulated (µl/min) (Mean ± SD) | Change (%) | |
| 1 | 56.6 ± 4.4 | 71.6 ± 10.0 | +26.5 (p = 0.15) |
| 2 | 42.4 ± 6.5 | 57.3 ± 5.4 | +35.1 (p = 0.03) |
| 3 | 41.8 ± 1.0 | 46.8 ± 1.7 | +12.0 (p = 0.18) |
| Group Overall | 46.9 ± 8.4 | 57.3 ± 13.5 | +22.2 (p = 0.002) |
P values are in comparison with the base line. Blood flow shown is the total venous flow.
Change in Total Vessel Cross-Sectional Area with Flicker Stimulation
| Subject | Total Venous Cross Section | Total Arterial Cross Section | ||||
|---|---|---|---|---|---|---|
| Base line (mm2) (Mean ± SD) | Stimulated (mm2) (Mean ± SD) | Change (%) | Baseline (mm2) (Mean ± SD) | Stimulated (mm2) (Mean ± SD) | Change (%) | |
| 1 | 0.065 ± 0.001 | 0.074 ± 0.003 | +13.8 (p = 0.02) | 0.047 ± 0.001 | 0.048 ± 0.001 | +2.1 (p = 0.13) |
| 2 | 0.048 ± 0.001 | 0.056 ± 0.001 | +16.7 (p<0.01) | 0.030 ± 0.001 | 0.032 ± 0.001 | +6.7 (p = 0.12) |
| 3 | 0.047 ± 0.001 | 0.049 ± 0.001 | +4.3 (p = 0.02) | 0.032 ± 0.002 | 0.033 ± 0.001 | +3.1 (p = 0.97) |
| Group Overall | 0.053 ± 0.010 | 0.059 ± 0.013 | +11.3 (p<0.001) | 0.036 ± 0.009 | 0.037 ± 0.010 | +2.7 (p = 0.28) |
Change in Average Flow Velocity with Flicker Stimulation.
| Average Venous Velocity | Average Arterial Velocity | ||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Subject | Base line (mm/s) (Mean ± SD) | Stimulated (mm/s) (Mean ± SD) | Change (%) | Baseline (mm/s) (Mean ± SD) | Stimulated (mm/s) (Mean ± SD) | Change (%) | |||||||
| 1 | 14.6 ± 1.3 | 16.0 ± 1.7 | +9.8 (p = 0.39) | 20.0 ± 1.3 | 24.6 ± 3.4 | +23.0 (p = 0.18) | |||||||
| 2 | 14.7 ± 1.9 | 16.8 ± 1.4 | +14.3 (p = 0.16) | 23.9 ± 4.3 | 29.8 ± 2.5 | +24.7 (p = 0.06) | |||||||
| 3 | 14.9 ± 0.6 | 15.3 ± 0.5 | +2.7 (p = 0.51) | 21.5 ± 0.8 | 23.0 ± 1.6 | +7.0 (p = 0.26) | |||||||
| Group Overall | 14.7 ± 0.2 | 16.0 ± 0.8 | +8.8 (p = 0.046) | 21.8 ± 1.9 | 25.8 ± 3.6 | +18.3 (p = 0.004) | |||||||
Arterial velocity is calculated by dividing the flow by the total arterial area in each eye