Accuracy of ten isoflurane vaporisers in current clinical use.

J M Kelly1, K-L Kong.   

Abstract

There is a lack of independent data regarding the accuracy of isoflurane vaporisers in current clinical use. We studied the accuracy of 10 isoflurane vaporisers (six Penlon Sigma Delta, four Datex-Ohmeda Isotec 5) taken from our operating theatres and anaesthetic rooms. Isoflurane output was measured at a range of fresh gas flow rates using both 100% oxygen and a 1:2 oxygen:nitrous oxide mixture as carrier gas. Within a clinical range of fresh gas flow rates (1-6 l.min(-1)) the worst relative output deviation of each vaporiser was at least ±20%, with under-delivery of isoflurane being commoner than over-delivery. Relative output deviation ranged from -62% to +21%, which could lead to a risk of awareness or overdose. Despite the existence of an in-house calibration schedule, the vaporisers we found in current clinical use under- or occasionally over-delivered isoflurane to a dangerous and unpredictable extent.
© 2011 The Authors. Anaesthesia © 2011 The Association of Anaesthetists of Great Britain and Ireland.

Entities:  

Mesh:

Substances:

Year:  2011        PMID: 21749341     DOI: 10.1111/j.1365-2044.2011.06756.x

Source DB:  PubMed          Journal:  Anaesthesia        ISSN: 0003-2409            Impact factor:   6.955


  2 in total

1.  Use of an Integrated Low-Flow Anesthetic Vaporizer, Ventilator, and Physiological Monitoring System for Rodents.

Authors:  Krista Bigiarelli; Luke E Schepers; Arvin H Soepriatna; Dave FitzMiller; Craig J Goergen
Journal:  J Vis Exp       Date:  2020-07-09       Impact factor: 1.355

2.  Use of a Low-flow Digital Anesthesia System for Mice and Rats.

Authors:  Amelia R Adelsperger; Krista J Bigiarelli-Nogas; Irina Toore; Craig J Goergen
Journal:  J Vis Exp       Date:  2016-09-07       Impact factor: 1.355

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.