Literature DB >> 21748725

Flushing of intravascular access devices (IVADs) - efficacy of pulsed and continuous infusions.

Gérard Guiffant1, Jean Jacques Durussel, Jacques Merckx, Patrice Flaud, Jean Pierre Vigier, Patrice Mousset.   

Abstract

PURPOSE: To compare, under controlled conditions similar to practical situations, the efficacy of pulsed and continuous infusion for flushing IVADs. For both of them different flow rates, flushing volumes, and times were tested.
METHODS: The PU catheter lumens were filled with a mixture of fibronectin and bovine albumin to simulate physiologic protein depot. Flushing was performed with normal saline. An adapted pump controlled the flow rates. Efficacy was measured by the amount of albumin recovered from the lumen of the tested devices. Flow rate, volumes, and times tested were based upon values reported and/or measured in nursing practices. We compared: (A) single 10 mL bolus and 6 flushing times, (B) continuous infusion of 500 mL in 24 hours, (C) 10 successive boluses 1 mL flushed in 0.5 s each and 6 different time intervals between each bolus. Statistics were performed using the Mann and Whitney U test.
RESULTS: In group A, the maximum percentage of recovered protein (79.1% was achieved with the 10 mL bolus flushed in 2.5. In group B, 77% of protein was recovered at 24h. In group C: maximum efficacy (90 +/- 3%) was obtained when the time interval between 2 boluses was 0.4 s.
CONCLUSIONS: We conclude that hydrodynamics has a determinant effect on the efficacy and that the adjunction of an intermittent component in the flow increases it. Flow type and the time interval between 2 boluses are the 2 critical variables.

Entities:  

Mesh:

Substances:

Year:  2012        PMID: 21748725     DOI: 10.5301/JVA.2011.8487

Source DB:  PubMed          Journal:  J Vasc Access        ISSN: 1129-7298            Impact factor:   2.283


  14 in total

1.  Blood Contamination of the Small Bore Peripheral Intravenous Catheter in Neonates.

Authors:  Vani Krishnamurthy; Srinivasa Murthy Doreswamy; Sushma Krishnagowda
Journal:  J Clin Diagn Res       Date:  2016-11-01

2.  Flushing ports of totally implantable venous access devices, and impact of the Huber point needle bevel orientation: experimental tests and numerical computation.

Authors:  Gérard Guiffant; Jean Jacques Durussel; Patrice Flaud; Jean Pierre Vigier; Jacques Merckx
Journal:  Med Devices (Auckl)       Date:  2012-04-16

Review 3.  Flushing and Locking of Venous Catheters: Available Evidence and Evidence Deficit.

Authors:  Godelieve Alice Goossens
Journal:  Nurs Res Pract       Date:  2015-05-14

4.  Pulsative flushing as a strategy to prevent bacterial colonization of vascular access devices.

Authors:  Agnès Ferroni; Florian Gaudin; Gérard Guiffant; Patrice Flaud; Jean-Jacques Durussel; Philippe Descamps; Patrick Berche; Xavier Nassif; Jacques Merckx
Journal:  Med Devices (Auckl)       Date:  2014-11-07

5.  Impact of the shape of the implantable ports on their efficiency of flow (injection and flushing).

Authors:  Gérard Guiffant; Patrice Flaud; Jean Jacques Durussel; Jacques Merckx
Journal:  Med Devices (Auckl)       Date:  2014-09-17

6.  The effect of different flushing methods in a short peripheral catheter.

Authors:  Cuiling Tong; Xiaochun Peng; Hong Hu; Zongwen Wang; Hong Zhou
Journal:  Acta Cir Bras       Date:  2019-10-14       Impact factor: 1.388

7.  Quantitative assessment of required separator fluid volume in multi-infusion settings.

Authors:  Frank Doesburg; Daniek Middendorp; Willem Dieperink; Wouter Bult; Maarten W Nijsten; Daan J Touw
Journal:  J Vasc Access       Date:  2020-05-04       Impact factor: 2.283

8.  Comparing the Effects of Pulsatile and Continuous Flushing on Time and Type of Peripheral Intravenous Catheters Patency: A Randomized Clinical Trial.

Authors:  Seyed Javad Hosseini; Fereshteh Eidy; Majid Kianmehr; Ali Asghar Firouzian; Fatemeh Hajiabadi; Mahmoud Marhamati; Mahbobeh Firooz
Journal:  J Caring Sci       Date:  2021-05-24

9.  Power port contrast medium flushing and trapping: impact of temperature, an in vitro experimental study.

Authors:  Gérard Guiffant; Jean Jacques Durussel; Patrice Flaud; Laurent Royon; Pierre Yves Marcy; Jacques Merckx
Journal:  Med Devices (Auckl)       Date:  2013-09-03

10.  Recommended irrigation volume for an intravenous port: Ex vivo simulation study.

Authors:  Ching-Yang Wu; Chia-Hui Cheng; Jui-Ying Fu; Yen Chu; Ching-Feng Wu; Chien-Hung Chiu; Po-Jen Ko; Yun-Hen Liu
Journal:  PLoS One       Date:  2018-08-14       Impact factor: 3.240