Literature DB >> 21737601

Evaluation of the left atrial appendage with real-time 3-dimensional transesophageal echocardiography: implications for catheter-based left atrial appendage closure.

Gaetano Nucifora1, Francesco F Faletra, François Regoli, Elena Pasotti, Giovanni Pedrazzini, Tiziano Moccetti, Angelo Auricchio.   

Abstract

BACKGROUND: Precise knowledge of left atrial appendage (LAA) orifice size is crucial for correct sizing of LAA closure devices. The aim of the present study was to determine the performance of real-time 3D transesophageal echocardiography (RT3DTEE) for LAA orifice size assessment, compared with 2D transesophageal echocardiography (2DTEE), and to investigate the impact of atrial fibrillation (AF) on LAA orifice size. METHODS AND
RESULTS: One hundred thirty-seven patients (38 control subjects, 31 with paroxysmal AF, 38 with persistent AF and 30 with permanent AF) underwent 2DTEE and RT3DTEE. Both techniques were used to measure LAA orifice area. Clinically-indicated 64-slice computed tomography (CT) was used as reference technique in 46 patients. Two-dimensional TEE underestimated LAA orifice area, compared with RT3DTEE (1.99±0.94 cm(2) versus 3.05±1.27 cm(2); P<0.001). RT3DTEE showed higher correlation with CT for the assessment of LAA orifice area, compared with 2DTEE (r=0.92; 95% confidence interval, 0.85 to 0.95, versus r=0.72; 95% confidence interval, 0.54 to 0.83, respectively). At Bland-Altman analysis, RT3DTEE and 2DTEE underestimated LAA orifice area, compared with CT. However, RT3DTEE showed smaller bias (0.07 cm(2) versus 0.72 cm(2)) and narrower limits of agreement (-0.71 to 0.85 cm(2) versus -0.58 to 2.02 cm(2)) with CT, compared with 2DTEE. Among AF patients, a progressive increase in RT3DTEE-derived LAA orifice area was observed with increasing frequency of AF (P<0.001). At multivariate analysis, AF and left atrial volume index (P<0.001 for both) were independently associated with RT3DTEE-derived LAA orifice area.
CONCLUSIONS: RT3DTEE is more accurate than 2DTEE for the assessment of LAA orifice size. A progressive increase in LAA orifice area is observed with increasing frequency of AF.

Entities:  

Mesh:

Year:  2011        PMID: 21737601     DOI: 10.1161/CIRCIMAGING.111.963892

Source DB:  PubMed          Journal:  Circ Cardiovasc Imaging        ISSN: 1941-9651            Impact factor:   7.792


  51 in total

1.  Initial clinical experience using the EchoNavigator(®)-system during structural heart disease interventions.

Authors:  Jan Balzer; Tobias Zeus; Katharina Hellhammer; Verena Veulemans; Silke Eschenhagen; Eva Kehmeier; Christian Meyer; Tienush Rassaf; Malte Kelm
Journal:  World J Cardiol       Date:  2015-09-26

2.  Interventional left atrial appendage occlusion: added value of 3D transesophageal echocardiography for device sizing.

Authors:  Björn Goebel; Stephanie Wieg; Ali Hamadanchi; Sylvia Otto; Christian Jung; Daniel Kretzschmar; Hans R Figulla; P Christian Schulze; Tudor C Poerner
Journal:  Int J Cardiovasc Imaging       Date:  2016-06-06       Impact factor: 2.357

Review 3.  Interventional imaging: the role of echocardiography.

Authors:  Dimitrios Maragiannis; Stephen H Little
Journal:  Methodist Debakey Cardiovasc J       Date:  2014 Jul-Sep

Review 4.  An update on intraoperative three-dimensional transesophageal echocardiography.

Authors:  Lisa Qia Rong
Journal:  J Thorac Dis       Date:  2017-04       Impact factor: 2.895

5.  3D-printed heart model to guide LAA closure: useful in clinical practice?

Authors:  Anne-Lise Hachulla; Stéphane Noble; Gabriel Guglielmi; Daniel Agulleiro; Hajo Müller; Jean-Paul Vallée
Journal:  Eur Radiol       Date:  2018-06-14       Impact factor: 5.315

Review 6.  [Imaging in structural heart disease : Impact on interventional therapy].

Authors:  A Schmermund; J Eckert; S N Schelle; H Eggebrecht
Journal:  Herz       Date:  2016-11       Impact factor: 1.443

Review 7.  Left Atrial Occlusion Device Implantation: the Role of the Echocardiographer.

Authors:  David Altszuler; Alan F Vainrib; Daniel G Bamira; Ricardo J Benenstein; Anthony Aizer; Larry A Chinitz; Muhamed Saric
Journal:  Curr Cardiol Rep       Date:  2019-06-10       Impact factor: 2.931

8.  Evaluation of left atrial appendage function and thrombi in patients with atrial fibrillation: from transthoracic to real time 3D transesophageal echocardiography.

Authors:  Ilaria Dentamaro; Domenico Vestito; Ennio Michelotto; Delia De Santis; Vittoria Ostuni; Christian Cadeddu; Paolo Colonna
Journal:  Int J Cardiovasc Imaging       Date:  2016-11-17       Impact factor: 2.357

Review 9.  [Echocardiography during transcatheter interventions. New developments].

Authors:  J Balzer
Journal:  Herz       Date:  2013-02       Impact factor: 1.443

10.  Implantation of the AMPLATZER™ Cardiac Plug: tips and tricks.

Authors:  Jai-Wun Park
Journal:  Herzschrittmacherther Elektrophysiol       Date:  2013-03-12
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.