Dosimetry using Gafchromic XR-RV2 radiochromic films in interventional radiology.

A Neocleous1, E Yakoumakis, G Gialousis, A Dimitriadis, N Yakoumakis, E Georgiou.   

Abstract

Patient dose measurements of local entrance dose to the skin have been carried out using radiochromic film (Gafchromic XR-RV2) in a sample of interventional procedures. The major aim of the work was to measure patient entrance dose from such examinations using Gafchromic XR-RV2. Forty-five various interventional procedures (including nefrostomies and urinary stenting, biliary stenting and percutaneous transhepatic biliary drainage (PTBD) and aorta stent grafting) were evaluated. Maximum entrance doses were 537 ± 119 mGy in nephrostomies, 943 ± 631 mGy in biliary stenting and PTBD and 2425 ± 569 mGy in aorta stent grafting. Results indicate that all patients undergoing aorta stent grafting received skin dose above 1500 mGy, which means that there is an increasing potential to suffer radiation-induced skin injuries. The film provides dose mapping, the position of the skin area with highest dose and can be used for immediate qualitative and as well as for quantitative assessment of patient skin dose.

Entities:  

Mesh:

Year:  2011        PMID: 21733864     DOI: 10.1093/rpd/ncr272

Source DB:  PubMed          Journal:  Radiat Prot Dosimetry        ISSN: 0144-8420            Impact factor:   0.972


  1 in total

1.  Dose evaluation for skin and organ in hepatocellular carcinoma during angiographic procedure.

Authors:  Daniela D'Alessio; Claudia Giliberti; Antonella Soriani; Livio Carpanese; Giuseppe Pizzi; Giulio Vallati; Lidia Strigari
Journal:  J Exp Clin Cancer Res       Date:  2013-10-25
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.