| Literature DB >> 21733858 |
L Gavagan1, J Ti, J Thornton.
Abstract
Reports in the literature of radiation-induced hair loss are becoming increasingly common. This work describes a retrospective dose study of patients (n = 958) undergoing diagnostic (primarily cerebral angiograms) and therapeutic (primarily cerebral embolisation) procedures in a neuro-interventional suite. A comparison of patient doses as dose area product (DAP) readings from a single-plane image intensifier system (mean DAP value of 8772 cGy cm²) were compared with patient doses from a flat panel biplane system (mean DAP value of 7855 cGy cm²). Over 80 % of patients requiring neuro-interventional procedures were found to undergo two procedures or more. An estimated 7 % of therapeutic procedures were found to reach the International Commission on Radiological Protection threshold for temporary epilation.Entities:
Mesh:
Year: 2011 PMID: 21733858 DOI: 10.1093/rpd/ncr270