| Literature DB >> 21726760 |
Xavier Cetó1, Francisco Céspedes, Josefina Capdevila, Manel del Valle.
Abstract
A new amperometric bienzymatic biosensor for gluconic acid based on the coimmobilization of gluconokinase (EC 2.7.1.12) and phosphogluconate dehydrogenase (EC 1.1.1.44) by polysulfone membrane entrapment onto the surface of a graphite-epoxy composite is reported. This biosensor represents an alternative to gluconate dehydrogenase (EC 1.1.99.3) based methods, which is no longer commercially available. Measurements were done at an applied potential of +0.800 V, room temperature and phosphate buffer pH 7.50; obtaining a linear response range for gluconic acid extended from 7.0 × 10(-6) to 2.5 × 10(-4)M. Constructed biosensors showed good reproducibility for calibrations using different electrodes (RSD of 1.74%). Finally, biosensor was applied to real wine samples, and the results obtained were validated by comparison with those provided by a reference laboratory. Good correlation was found when the biosensor results were plotted vs. the reference values (slope=1.03 ± 0.04, intercept=0.01 ± 0.02, r(2)=0.995).Entities:
Mesh:
Substances:
Year: 2011 PMID: 21726760 DOI: 10.1016/j.talanta.2011.05.047
Source DB: PubMed Journal: Talanta ISSN: 0039-9140 Impact factor: 6.057