Murtaza Hyder1, Lidia Luque Aguilar, Johan Genberg, Margareta Sandahl, Clas Wesén, Jan Åke Jönsson.
Abstract
A method for determination of polycyclic aromatic hydrocarbons (PAHs) from aerosols was developed. Instead of conventionally used non-polar or slightly polar phenylmethylpolysiloxane column a highly polar, highly substituted, cyanopropyl column (VF-23 MS) was used for separation of PAHs. Based on hollow fiber micro-porous membrane liquid-liquid extraction (HF-MMLLE) a method was developed for sample clean up and pretreatment. An enrichment factor of 617-1022 was obtained with extraction efficiency 10.2-18.9% for different PAHs analyzed in this study. The optimized method was successfully applied to aerosol samples and limits of detection between 1.2 pg m(-3) and 180 pg m(-3) was obtained. Almost all PAHs were found in most of the aerosol samples.Entities:
Year: 2011 PMID: 21726719 DOI: 10.1016/j.talanta.2011.04.080
Source DB: PubMed Journal: Talanta ISSN: 0039-9140 Impact factor: 6.057