| Literature DB >> 21719340 |
Annarita Cutrone1, Pier Nicola Sergi, Silvia Bossi, Silvestro Micera.
Abstract
In this paper a self-opening intrafascicular neural interface (SELINE) has been modeled using both a theoretical approach and a Finite Element (FE) analysis. This innovative self opening interface has several potential advantages such as: higher selectivity due to its three-dimensional structure and efficient anchorage system. Mechanical, structural and micro-technological issues have been considered to obtain an effective design of the electrode, as a feasibility study of the self-opening approach. A simple framework has been provided to model the insertion and partial retraction into peripheral nerves, resulting in the opening of wings. This integrated approach results in a rational procedure to optimize kinematics, geometry, and structural properties of peripheral interfaces. The design and feasibility study carried out in this work can potentially assure a correct behavior and dimensioning of the neural interface: in this way anomalous breakage should be avoided while mechanical and geometrical biocompatibility should increase.Entities:
Mesh:
Year: 2011 PMID: 21719340 DOI: 10.1016/j.medengphy.2011.06.001
Source DB: PubMed Journal: Med Eng Phys ISSN: 1350-4533 Impact factor: 2.242