Variables influencing the accuracy of 2-dimensional and real-time 3-dimensional echocardiography for assessment of small volumes, areas, and distances: an in vitro study using static tissue-mimicking phantoms.

Ulrike Herberg1, Manuel Brand, Christine Bernhardt, Hans Georg Trier, Johannes Breuer.   

Abstract

OBJECTIVES: The aim of this study was to assess the validity, accuracy, and reproducibility of real-time 3-dimensional (3D) echocardiography for small distances, areas, and volumes.
METHODS: Real-time 3D echocardiography using matrix technology was performed in small calibrated tissue-mimicking phantoms and compared with 2-dimensional (2D) echocardiography. In a systematic variation of variables on data acquisition and analysis including different 3D workstations (manual disk summation versus semiautomatic border detection), the relative contributions of sources of errors were determined. The clinical relevance of the in vitro findings was assessed in 5 neonates and infants.
RESULTS: Distance calculation was valid (mean relative error ± SD, -0.15% ± 1.2%). Underestimation of areas and volumes was significant for both 2D and 3D echocardiography (area: 2D, -7.0% ± 2.9%; 3D, -6.0% ± 2.8%; volume: 2D, -13.1% ± 4.5%; 3D, -6.7% ± 2.5%; P < .05). Adjustment of compression and gain on data acquisition (difference of the means: 2D, 11.6%; 3D, 17.9%), gain on postprocessing (3D, 3.4%), and the border detection algorithm on analysis (2D, 4.8%; 3D, 16.6%) had a highly significant effect on volume and area calculations (P < .001). In vivo, compression and gain on acquisition (3D, 19.1%) and the 3D workstation on analysis (3D, 22.2%) had a highly significant impact on left ventricular volumetry (P < .001).
CONCLUSIONS: Real-time 3D echocardiography is a reliable method for calculation of small distances, areas, and volumes comparable with the size of the neonatal and infant heart. Variables influencing boundary identification during image acquisition and analysis have a significant impact on 2D and 3D area and volume calculations. Standardized protocols are mandatory to avoid these sources of error in both clinical practice and research.

Entities:  

Mesh:

Year:  2011        PMID: 21705722     DOI: 10.7863/jum.2011.30.7.899

Source DB:  PubMed          Journal:  J Ultrasound Med        ISSN: 0278-4297            Impact factor:   2.153


  7 in total

1.  Multicenter study comparing shunt type in the norwood procedure for single-ventricle lesions: three-dimensional echocardiographic analysis.

Authors:  Gerald R Marx; Girish Shirali; Jami C Levine; Lin T Guey; James F Cnota; Jeanne M Baffa; William L Border; Steve Colan; Gregory Ensing; Mark K Friedberg; David J Goldberg; Salim F Idriss; J Blaine John; Wyman W Lai; Minmin Lu; Shaji C Menon; Richard G Ohye; David Saudek; Pierre C Wong; Gail D Pearson
Journal:  Circ Cardiovasc Imaging       Date:  2013-10-04       Impact factor: 7.792

2.  Ventricular pressure-volume loops obtained by 3D real-time echocardiography and mini pressure wire-a feasibility study.

Authors:  Ulrike Herberg; Eva Gatzweiler; Thomas Breuer; Johannes Breuer
Journal:  Clin Res Cardiol       Date:  2013-02-09       Impact factor: 5.460

3.  Comparison of stroke volumes assessed by three-dimensional echocardiography and transpulmonary thermodilution in a pediatric animal model.

Authors:  Katharina Linden; Dennis Ladage; Oliver Dewald; Eva Gatzweiler; Andrea Pieper; Matthias Seehase; Georg Daniel Duerr; Johannes Breuer; Ulrike Herberg
Journal:  J Clin Monit Comput       Date:  2016-02-17       Impact factor: 2.502

4.  An Optimistic View towards the Real Time 3D Echocardiography in Congenital Heart Disease: A Simple 'Crop Box' Should Give an Infinite Information in the Near Future!

Authors:  Jin-Hee Oh
Journal:  J Cardiovasc Ultrasound       Date:  2016-06-22

5.  Three-dimensional echocardiography in a dynamic heart phantom: comparison of five different methods to measure chamber volume using a commercially available software.

Authors:  Peter W Wood; Patrick H Gibson; Harald Becher
Journal:  Echo Res Pract       Date:  2014-10-09

6.  3D Real-Time Echocardiography Combined with Mini Pressure Wire Generate Reliable Pressure-Volume Loops in Small Hearts.

Authors:  Ulrike Herberg; Katharina Linden; Oliver Dewald; Eva Gatzweiler; Matthias Seehase; Georg Daniel Duerr; Jonas Dörner; Stephanie Kleppe; Dennis Ladage; Johannes Breuer
Journal:  PLoS One       Date:  2016-10-24       Impact factor: 3.240

7.  Assessment of pressure-volume relations in univentricular hearts: Comparison of obtainment by real-time 3D echocardiography and mini pressure-wire with conductance technology.

Authors:  Katharina Linden; Christian Winkler; Johannes Breuer; Ulrike Herberg
Journal:  PLoS One       Date:  2021-02-01       Impact factor: 3.240

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.