The use of a radiofrequency needle improves the safety and efficacy of transseptal puncture for atrial fibrillation ablation.

Roger A Winkle1, R Hardwin Mead, Gregory Engel, Rob A Patrawala.   

Abstract

BACKGROUND: Atrial fibrillation (AF) ablation requires transseptal puncture to gain entry to the left atrium (LA). On rare occasions, LA entry cannot be achieved or cardiac perforation results in pericardial tamponade.
OBJECTIVE: This study sought to compare a new radiofrequency (RF) transseptal needle with the standard needle.
METHODS: We evaluated 1,550 AF ablations in 1,167 patients. We compared 975 transseptal punctures done using a standard needle to 575 done using a new electrode-tipped needle attached to an RF perforation generator.
RESULTS: The rate of failure to cross the atrial septum was lower for the RF needle (1 of 575 [0.17%] vs. 12 of 975 [1.23%], P = .039) and there were fewer pericardial tamponades with the RF needle (0 of 575 [0.00%] vs. 9 of 975 [0.92%], P = .031). Multivariate analysis showed the RF needle use was the only variable associated with a lower incidence of tamponade (P = .04). Becasuse the RF needle was used later in our series, we examined our 975 standard needle punctures over time for evidence of improved operator experience that might explain the superior RF results. For the standard needle, there was no trend for improved septal crossing rates (P = .794) or fewer tamponades (P = .456) with more operator experience. Instrumentation time was shorter for the RF needle (27.1 ± 10.9 vs. 36.4 ± 17.7 minutes, P < .0001).
CONCLUSION: Our data suggest that the RF needle is superior to the standard transseptal needle. It results in shorter instrumentation times, a greater efficacy in transseptal crossing, and fewer episodes of pericardial tamponade.
Copyright © 2011 Heart Rhythm Society. Published by Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2011        PMID: 21699841     DOI: 10.1016/j.hrthm.2011.04.032

Source DB:  PubMed          Journal:  Heart Rhythm        ISSN: 1547-5271            Impact factor:   6.343


  19 in total

1.  Atrial fibrillation: RF needle improves septal puncture.

Authors:  Alexandra King
Journal:  Nat Rev Cardiol       Date:  2011-05-31       Impact factor: 32.419

2.  Guidewire electrosurgery-assisted trans-septal puncture.

Authors:  Jaffar M Khan; Toby Rogers; Marvin H Eng; Robert J Lederman; Adam B Greenbaum
Journal:  Catheter Cardiovasc Interv       Date:  2017-09-20       Impact factor: 2.692

Review 3.  Role of Intracardiac echocardiography in Atrial Fibrillation Ablation.

Authors:  Antonio Dello Russo; Eleonora Russo; Gaetano Fassini; Michela Casella; Ester Innocenti; Martina Zucchetti; Claudia Cefalu; Francesco Solimene; Gaetano Mottola; Daniele Colombo; Fabrizio Bologna; Benedetta Majocchi; Pasquale Santangeli; Stefania Riva; Luigi Di Biase; Cesare Fiorentini; Claudio Tondo
Journal:  J Atr Fibrillation       Date:  2013-04-06

4.  Persistence of an iatrogenic atrial septal defect after a second-generation cryoballoon ablation of atrial fibrillation.

Authors:  Tomonori Watanabe; Shinsuke Miyazaki; Takatsugu Kajiyama; Sadamitsu Ichijo; Takamitsu Takagi; Miyako Igarashi; Hiroaki Nakamura; Hiroshi Taniguchi; Hitoshi Hachiya; Yoshito Iesaka
Journal:  Heart Vessels       Date:  2018-03-17       Impact factor: 2.037

Review 5.  Visualization of catheter ablation for atrial fibrillation: Impact of devices and anatomy.

Authors:  Mark A Benscoter; Paul A Iaizzo
Journal:  World J Cardiol       Date:  2015-11-26

6.  Trends in atrial fibrillation ablation: have we maximized the current paradigms?

Authors:  Roger A Winkle; R Hardwin Mead; Gregory Engel; Melissa H Kong; Rob A Patrawala
Journal:  J Interv Card Electrophysiol       Date:  2012-02-28       Impact factor: 1.900

7.  Fluoroless left atrial access for radiofrequency and cryoballoon ablations using a novel radiofrequency transseptal wire.

Authors:  Hany Demo; Carla Aranda; Mansour Razminia
Journal:  J Interv Card Electrophysiol       Date:  2022-02-22       Impact factor: 1.759

8.  Radiofrequency ablation for paroxysmal atrial fibrillation in a patient with dextrocardia and interruption of the inferior vena cava: a case report.

Authors:  Xiaofeng Hu; Shaohui Wu; Mu Qin; Weifeng Jiang; Xu Liu
Journal:  Eur Heart J Case Rep       Date:  2021-05-19

9.  Physician-controlled costs: the choice of equipment used for atrial fibrillation ablation.

Authors:  Roger A Winkle; R Hardwin Mead; Gregory Engel; Melissa H Kong; Rob A Patrawala
Journal:  J Interv Card Electrophysiol       Date:  2013-03-14       Impact factor: 1.900

10.  Discontinuing anticoagulation following successful atrial fibrillation ablation in patients with prior strokes.

Authors:  Roger A Winkle; R Hardwin Mead; Gregory Engel; Melissa H Kong; Rob A Patrawala
Journal:  J Interv Card Electrophysiol       Date:  2013-10-08       Impact factor: 1.900

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.