Analytical evaluation of a high-sensitivity troponin T assay and its clinical assessment in acute coronary syndrome.

Camille Chenevier-Gobeaux1, Christophe Meune, Marie-Céline Blanc, Luc Cynober, Patrick Jaffray, Guillaume Lefevre.   

Abstract

BACKGROUND: The recently developed, highly sensitive cardiac troponin T (hs-cTnT) immunoassay improves the detection of acute myocardial infarction (AMI). However, this assay requires further analytical and clinical evaluation.
METHODS: Imprecision, linearity, limits of quantification and interferences were evaluated; hs-cTnT was compared with a conventional cardiac troponin I assay (cTnI), performed on an X-pand(®)HM, in a population of patients with suspected AMI. Finally, the 99th percentile cut-off point for a reference population was explored in 213 healthy control subjects.
RESULTS: Imprecision analysis demonstrated coefficients of variation (CVs) below 4%, linearity showed a 0.999 coefficient of correlation, with excellent recovery (99.9%) and a limit of quantification (10%CV) was found at 9.2 ng/L. A negative interference (>20%) with haemolysis was observed when supplemental haemoglobin was above 0.25 g/dL. Patients with suspected AMI more frequently displayed an increased hs-cTnT (83%) than an increased cTnI (55%, P < 0.01). Unstable angina was present in 63% of patients with an increased hs-cTnT associated with no increase in cTnI. The 99th percentile value for our reference population was 16.9 ng/L. In 213 healthy blood donors, hs-cTnT levels were significantly correlated with age (P < 0.0001), and were higher in men than in women (P < 0.0001).
CONCLUSIONS: The analytical performance of hs-cTnT complied with the international guidelines for AMI detection. Determining the degree of haemolysis in a sample is of paramount importance to the interpretation of hs-cTnT results. The 99th percentile value of our reference population was established.

Entities:  

Mesh:

Substances:

Year:  2011        PMID: 21697353     DOI: 10.1258/acb.2011.011019

Source DB:  PubMed          Journal:  Ann Clin Biochem        ISSN: 0004-5632            Impact factor:   2.057


  4 in total

Review 1.  Inosine and hypoxanthine as novel biomarkers for cardiac ischemia: from bench to point-of-care.

Authors:  Don E Farthing; Christine A Farthing; Lei Xi
Journal:  Exp Biol Med (Maywood)       Date:  2015-05-08

2.  Age- and sex-dependent upper reference limits for the high-sensitivity cardiac troponin T assay.

Authors:  M Odette Gore; Stephen L Seliger; Christopher R Defilippi; Vijay Nambi; Robert H Christenson; Ibrahim A Hashim; Ron C Hoogeveen; Colby R Ayers; Wensheng Sun; Darren K McGuire; Christie M Ballantyne; James A de Lemos
Journal:  J Am Coll Cardiol       Date:  2014-02-12       Impact factor: 24.094

3.  Age-sex distribution of patients with high-sensitivity troponin T levels below the 99th percentile.

Authors:  Jie-Yin Liu; Qiao-Wei Jia; Xiao-Ling Zang; Rong-Hu Wang; Chun-Jian Li; Lian-Sheng Wang; Wen-Zhu Ma; Zhi-Jian Yang; En-Zhi Jia
Journal:  Oncotarget       Date:  2017-08-18

4.  Biotin and high-sensitivity cardiac troponin T assay.

Authors:  Aurélien Schrapp; François Fraissinet; Charles Hervouet; Hélène Girot; Valéry Brunel
Journal:  Biochem Med (Zagreb)       Date:  2018-10-15       Impact factor: 2.313

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.