Abstract
This study was undertaken to evaluate matrix-assisted laser desorption ionization-time of flight mass spectrometry (MALDI-TOF MS) for the rapid identification of beta-hemolytic streptococci. We compared Bruker Biotyper 2.0 with Vitek2 coupled to the agglutination test. MALDI-TOF MS analysis of 386 beta-hemolytic streptococcal isolates yielded high-confidence identification to the species level for all 386 isolates. The Vitek2 gave high-confidence identification to the species level for 88% of Streptococcus agalactiae isolates (n = 269/306), 92% of Streptococcus pyogenes isolates (n = 48/52), and 39% of isolates of Streptococcus dysgalactiae serogroups C and G (n = 11/28).Entities:
Mesh:
Year: 2011 PMID: 21697322 PMCID: PMC3147758 DOI: 10.1128/JCM.00240-11
Source DB: PubMed Journal: J Clin Microbiol ISSN: 0095-1137 Impact factor: 5.948