Uncalibrated arterial pressure waveform analysis for cardiac output monitoring is biased by low peripheral resistance in patients with intracranial haemorrhage.

E K Junttila1, J K Koskenkari, P P Ohtonen, T I Ala-Kokko.   

Abstract

BACKGROUND: Cardiac output (CO) monitoring by uncalibrated arterial pressure waveform analysis (APCO) using the FloTrac/Vigileo™ is feasible in patients with intracranial haemorrhage, but the results of validation studies are contradictory. The aim of the present study was to analyse the clinical agreement between the intermittent bolus thermodilution technique (TDCO) and APCO in patients with non-traumatic intracranial haemorrhage.
METHODS: This was a prospective observational clinical study in a university level intensive care unit. We studied patients who underwent CO monitoring according to clinical indications using TDCO. Simultaneously, APCO was applied using the radial arterial pressure curve. The difference in CO values measured by APCO with a mid-chest calibration level was compared with a calibration level at the angle of the eye.
RESULTS: A total of 407 data pairs from 16 patients were obtained. The mean CO(TDCO) was 7.6 litre min(-1) and CO(APCO) was 6.0 litre min(-1), with a bias corrected for repeated measures of 1.5 litre min(-1) and 95% limits of agreement of -2.4 to 5.4 litre min(-1). The percentage error was 58%. The increasing bias correlated with low peripheral resistance (ρ=-0.53, P=0.036). The calibration level at the patient's eye angle did not affect CO values (median bias 0 litre min(-1) with 25th-75th percentile -0.1 to 0.2 litre min(-1)).
CONCLUSIONS: The second generation of FloTrac(®)/Vigileo(®) monitoring system underestimates the TDCO in patients with non-traumatic intracranial haemorrhage. The bias correlates with measured systemic vascular resistance. The upper calibration level does not affect the results.

Entities:  

Mesh:

Year:  2011        PMID: 21693470     DOI: 10.1093/bja/aer170

Source DB:  PubMed          Journal:  Br J Anaesth        ISSN: 0007-0912            Impact factor:   9.166


  6 in total

1.  The impact of systemic vascular resistance on the accuracy of the FloTrac/Vigileo™ system in the perioperative period of cardiac surgery: a prospective observational comparison study.

Authors:  Yohei Sotomi; Katsuomi Iwakura; Yoshiharu Higuchi; Kazuo Abe; Junko Yoshida; Takafumi Masai; Kenshi Fujii
Journal:  J Clin Monit Comput       Date:  2013-06-08       Impact factor: 2.502

2.  Automated stroke volume and pulse pressure variations predict fluid responsiveness in mechanically ventilated patients with obstructive jaundice.

Authors:  Feng Zhao; Peng Wang; Shujun Pei; Weidong Mi; Qiang Fu
Journal:  Int J Clin Exp Med       Date:  2015-11-15

Review 3.  Advanced monitoring of systemic hemodynamics in critically ill patients with acute brain injury.

Authors:  Fabio Silvio Taccone; Giuseppe Citerio
Journal:  Neurocrit Care       Date:  2014-12       Impact factor: 3.210

4.  Bioreactance and fourth-generation pulse contour methods in monitoring cardiac index during off-pump coronary artery bypass surgery.

Authors:  Laura Anneli Ylikauma; Pasi Petteri Ohtonen; Tiina Maria Erkinaro; Merja Annika Vakkala; Janne Henrik Liisanantti; Jari Uolevi Satta; Tatu Sakari Juvonen; Timo Ilari Kaakinen
Journal:  J Clin Monit Comput       Date:  2021-05-26       Impact factor: 1.977

5.  Accuracy of Cardiac Output by Nine Different Pulse Contour Algorithms in Cardiac Surgery Patients: A Comparison with Transpulmonary Thermodilution.

Authors:  Ole Broch; Berthold Bein; Matthias Gruenewald; Sarah Masing; Katharina Huenges; Assad Haneya; Markus Steinfath; Jochen Renner
Journal:  Biomed Res Int       Date:  2016-12-28       Impact factor: 3.411

6.  Accuracy of an autocalibrated pulse contour analysis in cardiac surgery patients: a bi-center clinical trial.

Authors:  Ole Broch; Jose Carbonell; Carlos Ferrando; Malte Metzner; Arne Carstens; Martin Albrecht; Matthias Gruenewald; Jan Höcker; Marina Soro; Markus Steinfath; Jochen Renner; Berthold Bein
Journal:  BMC Anesthesiol       Date:  2015-11-26       Impact factor: 2.217

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.