Accidental carotid artery cannulation detected by bedside ultrasound.

Kris Chiles1, Arun Nagdev.   

Abstract

This report highlights the importance of using bedside ultrasound in the emergency department to confirm guide-wire placement when performing central venous catheter placement prior to dilating and cannulating the vessel.

Entities:  

Year:  2011        PMID: 21691480      PMCID: PMC3088383     

Source DB:  PubMed          Journal:  West J Emerg Med        ISSN: 1936-900X


A 73-year-old female was brought to the emergency department for altered level of consciousness. Her blood pressure was 81/50 mmHg, which did not respond to fluid hydration. The right internal jugular (IJ) vein was cannulated with ultrasound guidance for vasoactive administration. A flash-back of dark non-pulsatile blood was obtained before vessel dilation and catheter placement. A portable chest radiograph was performed and read by the staff radiologist as “left IJ catheter is seen overlying the mediastinum likely within the left innominate vein” (Figure 1). Prior to the use of the catheter, the nursing team noted abnormal pulsations from the distal port of central venous catheter (CVC), prompting a bedside ultrasound by the clinician to ensure proper placement (Figure 2). Differentiation of the carotid artery and IJ vein was difficult secondary to the patient’s severe volume depletion, but doppler showed pulsatile flow in the cannulated vessel confirming inadvertent arterial placement.
Figure 1.

Chest radiograph shows central venous catheter overlying left internal jugular vein traversing the mediastinum.

Figure 2.

Ultrasound visualization of catheter within the lumen of the left carotid artery with conformation using color flow dropper.

This report highlights the use of using ultrasound to confirm proper placement of a central venous catheters. The IJ vein overlies the carotid artery, making chest radiograph an unreliable test to ensure venous placement. Central venous pressure (CVP) monitoring can confirm venous cannulation, but can be difficult to obtain in many emergency departments. Classic teaching of looking for dark non-pulsatile blood can be inaccurate in hypotensive and hypoxic patients. Also when a collapsible vein overlies the carotid artery, a through puncture can occur.1 A prior study demonstrated the use of guidewire visualization with bedside ultrasound as a useful technique to confirm venous placement prior to dilation.2 We feel that with the fair amount of literature demonstrating improved safety of ultrasound-guidance in CVC, confirming venous placement of either the guidewire or catheter can be a simple addition in hypotensive patients or those that uncertainty exists regarding arterial cannulation.
  2 in total

1.  Ultrasound detection of guidewire position during central venous catheterization.

Authors:  Michael B Stone; Arun Nagdev; Michael C Murphy; Craig A Sisson
Journal:  Am J Emerg Med       Date:  2010-01       Impact factor: 2.469

2.  An unseen danger: frequency of posterior vessel wall penetration by needles during attempts to place internal jugular vein central catheters using ultrasound guidance.

Authors:  Michael Blaivas; Srikar Adhikari
Journal:  Crit Care Med       Date:  2009-08       Impact factor: 7.598

  2 in total
  1 in total

Review 1.  Clinician-performed ultrasound in hemodynamic and cardiac assessment: a synopsis of current indications and limitations.

Authors:  N Kelly; R Esteve; T J Papadimos; R P Sharpe; S A Keeney; R DeQuevedo; M Portner; D P Bahner; S P Stawicki
Journal:  Eur J Trauma Emerg Surg       Date:  2015-01-08       Impact factor: 3.693

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.