| Literature DB >> 21687539 |
Ming-Yu Chao1, Kung-Tien Liu, Yi-Chih Hsia, Mei-Hsiu Liao, Lie-Hang Shen.
Abstract
Technetium-99m ethyl cysteinate dimer (Tc-99m-ECD) is an essential imaging agent used in evaluating the regional cerebral blood flow in patients with cerebrovascular diseases. Determination of active pharmaceutical ingredient, that is, L-Cysteine, N, N'-1,2-ethanediylbis-, diethyl ester, dihydrochloride (ECD) in ECD Kit is a relevant requirement for the pharmaceutical quality control in processes of mass fabrication. We here presented a direct solid sample determination method of ECD in ECD Kit without sample dissolution to avoid the rapid degradation of ECD. An elemental analyzer equipped with a nondispersive infrared detector and a calibration curve of coal standard was used for the quantitation of sulfur in ECD Kit. No significant matrix effect was found. The peak area of coal standard against the amount of sulfur was linear over the range of 0.03-0.10 mg, with a correlation coefficient (r) of 0.9993. Method validation parameters were achieved to demonstrate the potential of this method.Entities:
Mesh:
Substances:
Year: 2011 PMID: 21687539 PMCID: PMC3114541 DOI: 10.1155/2011/196238
Source DB: PubMed Journal: J Biomed Biotechnol ISSN: 1110-7243
Optimized parameters of elemental analyzer for quantitation of ECD in ECD Kit.
| Parameters | |
|---|---|
| Operation mode | CHNS |
| Combustion temperature (°C) | 1150 |
| Reduction temperature (°C) | 900 |
| Flush gas/time (sec) | He/10 |
| O2 dosing time (sec) | 120 |
| Column standby temperature (°C) | |
| CO2 | Ambient temperature |
| H2O | Ambient temperature |
| SO2 | 140 |
| Column desorption temperature (°C) | |
| CO2 | 240 |
| H2O | 150 |
| SO2 | 220 |
| Carrier gas/Flow rate (mL/min)(1) | He/230 |
| Flow rate of O2 (mL/min) | 15 |
| Flow rate of O2 during combustion (mL/min) | 30–35 |
(1)Same as the mass flow control (MFC) TCD flowing gas and flow rate.
Figure 1Typical elemental analyzer chromatogram of ECD in ECD Kit.
Preparation and composition of ECD calibration standards, blank, and quality control samples.
| Sample | WECD (1) (mg/vial) | WKit (2) (mg/vial) | WS (3) (mg/vial) | %WS (4) (%, w/w) |
|---|---|---|---|---|
| Kit blank (BkKit) | 0.00 | 25.61 | 0.00 | 0.00 |
| ECD Calibration standards | ||||
| StdECD no. 1 | 0.78 | 26.27 | 0.123 | 0.47 |
| StdECD no. 2 | 0.89 | 26.00 | 0.140 | 0.54 |
| StdECD no. 3 | 0.97 | 26.10 | 0.152 | 0.58 |
| StdECD no. 4 | 1.07 | 26.04 | 0.168 | 0.65 |
| ECD QC samples (QCECD) | ||||
| QC-L | 0.23 | 7.27 | 0.036 | 0.48 |
| QC-M | 0.27 | 7.23 | 0.042 | 0.57 |
| QC-H | 0.31 | 7.19 | 0.049 | 0.65 |
(1)Nominal weight of ECD in ECD Kit.
(2)Total weight of ECD Kit.
(3)Nominal weight of sulfur in ECD Kit.
(4)Percentage of sulfur (%, w/w) in ECD Kit.
Precision and accuracy in the analysis of QC samples and ECD in ECD Kit.
| Day | Standard curve(1) | Sulfur weight (%)(2) | Recovery yield (%) | |||
|---|---|---|---|---|---|---|
| Dynamic range of sulfur (mg) | Linear least squares regression equation | Correlation coefficient ( | QCcoal | QCcoal | QCECD (3) | |
| 1 | 0.033 –0.104 | Y = 1.615 × 10−6X + 4.747 × 10−3 | 0.9993 | 3.13 ± 0.07 (2.25%) | 102.08 ± 2.29 | — |
| 1 | 0.031–0.105 | Y = 1.623 × 10−6X + 1.741 × 10−3 | 0.9989 | 3.10 ± 0.02 (0.60%) | 100.89 ± 0.60 | — |
| 2 | 0.034–0.107 | Y = 1.634 × 10−6X + 1.034 × 10−3 | 0.9994 | 3.08 ± 0.04 (1.21%) | 100.15 ± 1.21 | — |
| 3 | 0.033–0.106 | Y = 1.576 × 10−6X + 4.202 × 10−3 | 0.9996 | 3.18 | 103.79 | 102.78 |
| 100.00 | ||||||
| 102.08 | ||||||
(1)Standard curves of coal.
(2)Content percentage of sulfur in coal standard: 3.07% (w/w); data are expressed as average ± SD (%R.S.D.), n = 3.
(3)Purity of ECD: 97.53%; compositions of ECD QC samples (QC-L, QC-M, and QC-H) were shown in Table 2.
Stability study of QC samples analysis.
| Day | Standard curve | Recovery yield (%)(1) | |||
|---|---|---|---|---|---|
| Dynamic range of sulfur (mg) | Linear least squares regression equation | Correlation coefficient ( | QCcoal (2) | QCECD (3) | |
| 1 | 0.032–0.105 | Y = 1.627 × 10−6X + 4.149 × 10−3 | 0.9994 | 101.10 ± 0.94 | 96.02 ± 2.33 |
| 2 | 0.033–0.103 | Y = 1.629 × 10−6X + 2.390 × 10−3 | 0.9994 | 99.82 ± 0.90 | 98.48 ± 1.96 |
| 3 | 0.031–0.105 | Y = 1.609 × 10−6X + 2.608 × 10−3 | 0.9997 | 101.72 ± 2.00 | 102.31 ± 1.63 |
(1)Data are expressed as average ± SD, n = 3.
(2)QCcoal: 2.05 ± 0.05 mg of coal QC samples (S = 3.07%, w/w) were analyzed.
(3)QCECD: 7.52 ± 0.03 mg of ECD QC samples (ECD = 3.61%; S = 0.58%, w/w) were analyzed.
Robustness study in the analysis of ECD.
| Parameter | Standard curve of coal(1) | QCcoal (2) | |||
|---|---|---|---|---|---|
| Linear least squares regression equation | Correlation coefficient ( | Sulfur weight (%) | Recovery yield (%)(3) | ||
| Dosing time (sec) | 90 | Y = 1.544 × 10−6X + 3.394 × 10−3 | 0.9992 | 3.14 ± 0.06 | 102.36 ± 2.58 |
| 120 | Y = 1.615 × 10−6X + 4.747 × 10−3 | 0.9993 | 3.13 ± 0.07 | 101.91 ± 2.30 | |
| 150 | Y = 1.604 × 10−6X + 7.508 × 10−4 | 0.9999 | 3.05 ± 0.10 | 99.17 ± 2.84 | |
| Temperature of combustion tube (°C) | 1120 | Y = 1.605 × 10−6X + 8.215 × 10−4 | 0.9997 | 3.13 ± 0.06 | 102.01 ± 1.85 |
| 1150 | Y = 1.615 × 10−6X + 4.747 × 10−3 | 0.9993 | 3.13 ± 0.07 | 101.91 ± 2.30 | |
| 1180 | Y = 1.586 × 10−6X + 1.126 × 10−3 | 0.9985 | 3.03 ± 0.03 | 98.68 ± 1.14 | |
| Temperature of reduction tube (°C) | 850 | Y = 1.621 × 10−6X + 9.226 × 10−5 | 0.9997 | 3.12 ± 0.06 | 102.01 ± 1.75 |
| 900 | Y = 1.615 × 10−6X + 4.747 × 10−3 | 0.9993 | 3.13 ± 0.07 | 101.91 ± 2.30 | |
| 950 | Y = 1.649 × 10−6X − 1.288 × 10−3 | 0.9996 | 3.00 ± 0.02 | 97.97 ± 0.87 | |
(1)Standard curves were constructed by the coal concentration range of 1.01 to 3.49 mg.
(2)Data are expressed as average ± SD, n = 3.
(3)Recovery yield (%) = Sulfur(mg)exp/Sulfur(mg)nominal × 100%.