Inaccuracy of seven popular sphygmomanometers for home measurement of blood pressure.

E O'Brien1, F Mee, N Atkins, K O'Malley.   

Abstract

Seven models, available commercially for the self-measurement of blood pressure, were subjected to a validation procedure in which three devices of each model were tested by observers who were trained to a high standard of accuracy. The models were the Omron HEM-400C, the Philips HP5308, the Healthcheck 'Cuffless' CX-5 060020, the Nissei Analogue Monitor, the Philips HP5306/B, the Systema Dr MI-150 and the Fortec Dr MI-100. The validation programme had a number of unique features which included assessment of interdevice variability before and after 1 month of home use, and a new form of analysis, which we term 'clinical', based on the likely influence of three grades of device inaccuracy on patient management. In the main validation phase, one device of each model was compared with simultaneous measurements made by two 'blinded' observers using a standard mercury sphygmomanometer (PyMaH Corporation, New Jersey, USA) in the same arm in 85 subjects with a wide range of blood pressures. Three models (the Healthcheck 'Cuffless' CX-5 060020, the Systema Dr MI-150 and the Fortec Dr MI-100) failed the interdevice variability tests and did not reach the main validation test. Two models (the Omron HEM-400C and the Philips HP5306/B) failed on the criteria set down by the American National Standard for Electronic or Automated Sphygmomanometers, as well as the 'clinical' criteria. The remaining two models (the Nissei Analogue Monitor and the Philips HP5308) were acceptable for the measurement of systolic blood pressure by both methods of analysis but failed in the 'clinical' analysis for diastolic blood pressure. The mercury sphygmomanometer was comfortably within the criteria for both methods of analysis.

Entities:  

Mesh:

Year:  1990        PMID: 2168452     DOI: 10.1097/00004872-199007000-00005

Source DB:  PubMed          Journal:  J Hypertens        ISSN: 0263-6352            Impact factor:   4.844


  9 in total

1.  Replacing the mercury sphygmomanometer. Requires clinicians to demand better automated devices.

Authors:  E O'Brien
Journal:  BMJ       Date:  2000-03-25

2.  Measurement and interpretation of blood pressure.

Authors:  C D Goonasekera; M J Dillon
Journal:  Arch Dis Child       Date:  2000-03       Impact factor: 3.791

3.  Blood pressure measuring devices: recommendations of the European Society of Hypertension.

Authors:  E O'Brien; B Waeber; G Parati; J Staessen; M G Myers
Journal:  BMJ       Date:  2001-03-03

Review 4.  Importance of various methods of blood pressure measurement in clinical trials.

Authors:  P Palatini
Journal:  Curr Hypertens Rep       Date:  2000-08       Impact factor: 5.369

5.  Pressure for change: unresolved issues in blood pressure measurement.

Authors:  M Aylett
Journal:  Br J Gen Pract       Date:  1999-02       Impact factor: 5.386

6.  Insulin resistance and risk factors for cardiovascular disease in young adult survivors of childhood acute lymphoblastic leukemia.

Authors:  Kevin C Oeffinger; Beverley Adams-Huet; Ronald G Victor; Timothy S Church; Peter G Snell; Andrea L Dunn; Debra A Eshelman-Kent; Robert Ross; Peter M Janiszewski; Alicia J Turoff; Sandra Brooks; Gloria Lena Vega
Journal:  J Clin Oncol       Date:  2009-06-29       Impact factor: 44.544

7.  Association of African Ancestry With Electrocardiographic Voltage and Concentric Left Ventricular Hypertrophy: The Dallas Heart Study.

Authors:  Aya J Alame; Sonia Garg; Julia Kozlitina; Colby Ayers; Ronald M Peshock; Susan A Matulevicius; Mark H Drazner
Journal:  JAMA Cardiol       Date:  2018-12-01       Impact factor: 14.676

8.  A proposal for scientific validation of instruments for indirect blood pressure measurement at rest, during exercise, and in critical care.

Authors:  Y Iyriboz; C M Hearon
Journal:  J Clin Monit       Date:  1994-05

9.  The pursuit of accurate blood pressure measurement: A 35-year travail.

Authors:  Eoin O'Brien; George S Stergiou
Journal:  J Clin Hypertens (Greenwich)       Date:  2017-04-21       Impact factor: 3.738

  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.