Literature DB >> 21669960

Safety, efficacy, and performance of new discrimination algorithms to reduce inappropriate and unnecessary shocks: the PainFree SST clinical study design.

Angelo Auricchio1, Albert Meijer, Takashi Kurita, Edward Schloss, Karin Brinkman, Marieke Claessens-van Ooijen, Laurence Sterns.   

Abstract

INTRODUCTION: Implantable cardioverter defibrillator (ICD) shock therapy improves survival of patients at risk for sudden cardiac death. The high sensitivity of ICDs to detect tachycardia events is accompanied by reduced specificity resulting in inappropriate and unnecessary shocks. Up to 30% of ICD patients may experience inappropriate shocks, which are most commonly caused by lead noise, oversensing of T-waves, and supraventricular tachycardias. The new Protecta ICD and cardiac resynchronization therapy devices have been designed to minimize inappropriate and unnecessary shocks through novel SmartShock(TM) technology algorithms targeting these causes.
METHODS: The PainFree SST study is a prospective, multicentre clinical trial, which will be conducted in two consecutive phases. Phase I will assess safety and any delay that may arise in ventricular fibrillation (VF) arrhythmia detection time using new algorithms. Phase II will evaluate reduction of inappropriate and unnecessary shocks at 1 year of follow-up. Additional objectives will include Quality of Life, healthcare utilization, safety of extending the ventricular tachyarrhythmia/VF interval detection duration (18 out of 24 vs. 30 out of 40 intervals), and reasons for inappropriate shock. Up to 2000 subjects in 150 centres worldwide will be enrolled with a follow-up of at least 1 year. Subjects enrolled in Phase I will continue in Phase II of the study and data from all enrolled subjects will contribute to the analysis of Phase II objectives.
CONCLUSION: Inappropriate and unnecessary shock delivery remains a significant clinical issue for patients receiving device therapies, which has considerable consequences for patients and the healthcare system. The PainFree SST study will investigate the ability of new algorithms to reduce inappropriate shocks. Results from this study are expected in mid-2013.

Entities:  

Mesh:

Year:  2011        PMID: 21669960     DOI: 10.1093/europace/eur133

Source DB:  PubMed          Journal:  Europace        ISSN: 1099-5129            Impact factor:   5.214


  5 in total

1.  Reduction of the inappropriate ICD therapies by implementing a new fuzzy logic-based diagnostic algorithm.

Authors:  Michał Lewandowski; Andrzej Przybylski; Wiesław Kuźmicz; Hanna Szwed
Journal:  Ann Noninvasive Electrocardiol       Date:  2013-09       Impact factor: 1.468

2.  Predictors and clinical impact of inappropriate implantable cardioverter-defibrillator shocks in Korean patients.

Authors:  Jeong Hoon Yang; Kyeongmin Byeon; Hye Ran Yim; Jung Wae Park; Seung-Jung Park; June Huh; June Soo Kim; Young Keun On
Journal:  J Korean Med Sci       Date:  2012-05-26       Impact factor: 2.153

3.  Influence of myopotential interference on the Wavelet discrimination algorithm in implantable cardioverter-defibrillator.

Authors:  Kazuya Mizukami; Hisashi Yokoshiki; Hirofumi Mitsuyama; Masaya Watanabe; Taro Tenma; Rui Kamada; Masayuki Takahashi; Ryo Sasaki; Motoki Maeno; Hiroyuki Tsutsui
Journal:  J Arrhythm       Date:  2016-10-01

4.  Low rate of and rapid attention to inappropriate ICD shocks with remote device and rhythm monitoring: a qualitative study.

Authors:  Emma Sandgren; Cecilia Rorsman; Johan Engdahl; Nils Edvardsson
Journal:  Open Heart       Date:  2015-07-27

5.  State of the art of ICD programming: Lessons learned and future directions.

Authors:  M H Mastenbroek; S S Pedersen; H Versteeg; P A Doevendans; M Meine
Journal:  Neth Heart J       Date:  2014-10       Impact factor: 2.380

  5 in total