| Literature DB >> 21656566 |
Abstract
In this study, the reverse polarization method is implemented using transmit and receive arrays to improve the visibility of the interventional devices. Linearly polarized signal sources--inductively and receptively coupled radiofrequency coils--are used in the experimental setups to demonstrate the ability of the method to separate these sources from a forward polarized anatomy signal. Two different applications of the reverse polarization method are presented here: (a) catheter tracking and (b) fiducial marker visualization, in both of which transmit and receive arrays are used. The performance of the reverse polarization method was further tested with phantom and volunteer studies, and the results proved the feasibility of this method with transmit and receive arrays.Mesh:
Year: 2011 PMID: 21656566 DOI: 10.1002/mrm.23030
Source DB: PubMed Journal: Magn Reson Med ISSN: 0740-3194 Impact factor: 4.668