Literature DB >> 21647889

Cochlear reimplantation with same device: surgical and audiologic results.

Kim S van der Marel1, Jeroen J Briaire, Berit M Verbist, Raoul M S Joemai, Peter-Paul B M Boermans, Ferdinand A W Peek, Johan H M Frijns.   

Abstract

OBJECTIVES/HYPOTHESIS: To study to what extent it is possible to achieve identical insertion depths and to maintain the same performance after cochlear reimplantation. STUDY
DESIGN: Outcome research on a retrospective case series in a tertiary university referral center.
METHODS: Data were collected for 12 adults and three children who underwent reimplantation during the last 3 years with a new HiRes90K device with HiFocus 1J electrode owing to failure of the feed-through seal. Multislice computed tomography scans were used to compare positions of the original and newly placed electrode arrays. The speech-perception scores on a consonant-vowel-consonant word test before and after reimplantation were compared.
RESULTS: All reimplantations were successfully performed by two experienced cochlear implantation surgeons, and no complications were observed. Postoperative imaging showed that the average displacement of the new implant was only 0.59 mm. Reactivation of the implant gave immediate open set speech understanding in all patients, and speech perception rapidly returned to the previous level obtained with the original implant within weeks; it was even significantly better at the 3-month follow-up. No relation was found between changes in performance and the amount of displacement of the electrode array.
CONCLUSIONS: After cochlear reimplantation with the same device, electrode-array position can be accurately replicated and speech perception can be regained or even improved within weeks.
Copyright © 2011 The American Laryngological, Rhinological, and Otological Society, Inc.

Entities:  

Mesh:

Year:  2011        PMID: 21647889     DOI: 10.1002/lary.21722

Source DB:  PubMed          Journal:  Laryngoscope        ISSN: 0023-852X            Impact factor:   3.325


  3 in total

1.  Round window versus cochleostomy technique in cochlear implantation: histologic findings.

Authors:  Céline Richard; Jose N Fayad; Joni Doherty; Fred H Linthicum
Journal:  Otol Neurotol       Date:  2012-09       Impact factor: 2.311

2.  Audiological Outcomes and Associated Factors after Pediatric Cochlear Reimplantation.

Authors:  Fabian Blanc; Catherine Blanchet; Marielle Sicard; Fanny Merklen; Frederic Venail; Michel Mondain
Journal:  J Clin Med       Date:  2022-06-01       Impact factor: 4.964

3.  Electrode Afterload: A Valuable Technique in a Case of Short Electrode Insertion.

Authors:  C Riemann; S Scholz; H Sudhoff; I Todt
Journal:  Case Rep Otolaryngol       Date:  2020-02-13
  3 in total