Literature DB >> 21646659

Late cardiac perforation following percutaneous atrial septal defect closure using the Solysafe device.

Marcelo Cabrera1, Alejandro Contreras, Alejandro Peirone.   

Abstract

Percutaneous atrial septal defect closure has been demonstrated to be a safe and effective procedure. Many different devices have been utilized during the last decade with a low complication rate. We report a patient who presented with cardiac tamponade 6 months after ASD closure using the Solysafe device. Emergency surgery demonstrated a disarticulated device as well as a broken wire migrated into the pericardial space causing cardiac perforation and tamponade. To the best of our knowledge, this is the first report of late cardiac perforation following ASD closure using the Solysafe device.

Entities:  

Mesh:

Year:  2011        PMID: 21646659

Source DB:  PubMed          Journal:  J Invasive Cardiol        ISSN: 1042-3931            Impact factor:   2.022


  1 in total

1.  Mid-term results of Solysafe® septal occluder for percutaneous closure of secundum atrial septal defects.

Authors:  Murat Muhtar Yılmazer; Selman Vefa Yıldırım; Timur Meşe; Barış Güven; Savaş Demirpençe; Rahmi Özdemir; Taliha Öner; Vedide Tavlı
Journal:  Turk Gogus Kalp Damar Cerrahisi Derg       Date:  2018-01-09       Impact factor: 0.332