| Literature DB >> 21645713 |
Burkhard Horstkotte1, Carlos M Duarte, Víctor Cerdà.
Abstract
In this article, a simple, economic, and miniature flow analyzer for ammonium in seawater based on the solenoid micropumps is presented. A single reagent of sodium tetraborate, ortho-phthaldialdehyde (OPA), and sodium sulfite was used and optimized applying the modified SIMPLEX method. A special-made detection cell for fluorescence detection of the reaction product isoindol-1-sulfonat was made and combined with a commercial photomultiplier tube, a long-pass optical filter, and an UV-LED as excitation light source. A LOD down to 13 nmol/L was achieved. The fabrication and application of a miniature reaction coil heating device for reaction rate enhancement is further described. The system featured an injection frequency of 32 h(-1) at average standard deviation of 3%.Entities:
Mesh:
Substances:
Year: 2011 PMID: 21645713 DOI: 10.1016/j.talanta.2011.03.078
Source DB: PubMed Journal: Talanta ISSN: 0039-9140 Impact factor: 6.057