High-resolution mass spectrometry with a multiple pass quadrupole mass analyzer.

M H Amad1, R S Houk.   

Abstract

The peak shape narrows and the resolution improves if the ions are simply reflected back and forth through a conventional quadrupole mass analyzer. CO(+) and N(2)(+) at m/z = 28 are separated to 50% valley with half of the original signal remaining. These two ions can be resolved to baseline (m/Δm) = 5000 with 1% of the original signal remaining.

Entities:  

Year:  1998        PMID: 21644671     DOI: 10.1021/ac980505w

Source DB:  PubMed          Journal:  Anal Chem        ISSN: 0003-2700            Impact factor:   6.986


  1 in total

1.  Mass resolution of 11,000 to 22,000 with a multiple pass quadrupole mass analyzer