| Literature DB >> 21636260 |
Alessandra Bini1, Marcello Mascini, Marco Mascini, Anthony P F Turner.
Abstract
The possibility of introducing a computationally assisted method to study aptamer-protein interaction was evaluated with the aim of streamlining the screening and selection of new aptamers. Starting from information on the 15-mer (5'-GGTTGGTGTGGTTGG-3') thrombin binding aptamer (TBA), a library of mutated DNA sequences (994 elements) was generated and screened using shapegauss a shape-based scoring function from openeye software to generate computationally derived binding scores. The TBA and three other mutated oligonucleotides, selected on the basis of their binding score (best, medium, worst), were incorporated into surface plasmon resonance (SPR) biosensors. By reducing the ionic strength (binding buffer, 50 mM TrisHCl pH 7.4, 140 mM NaCl, 1mM MgCl₂, diluted 1:50) in order to match the simulated condition, the analytical performances of the four oligonucleotide sequences were compared using signal amplitude, sensitivity (slope), linearity (R²) and reproducibility (CVav %). The experimental results were in agreement with the simulation findings.Entities:
Mesh:
Substances:
Year: 2011 PMID: 21636260 DOI: 10.1016/j.bios.2011.04.053
Source DB: PubMed Journal: Biosens Bioelectron ISSN: 0956-5663 Impact factor: 10.618