Reliability and validity of the INFOOT three-dimensional foot digitizer for patients with rheumatoid arthritis.

Sophie De Mits1, Pascal Coorevits, Dirk De Clercq, Dirk Elewaut, James Woodburn, Philip Roosen.   

Abstract

BACKGROUND: Abnormal foot posture and deformities are identified as important features in rheumatoid arthritis. There is still no consensus regarding the optimum technique(s) for quantifying these features; hence, a foot digitizer might be used as an objective measurement tool. We sought to assess the validity and reliability of the INFOOT digitizer.
METHODS: To investigate the validity of the INFOOT digitizer compared with clinical measurements, we calculated Pearson correlation coefficients. To investigate the reliability of the INFOOT digitizer, we calculated intraclass correlation coefficients, SEMs, smallest detectable differences, and smallest detectable difference percentages.
RESULTS: Most of the 38 parameters showed good intraclass correlation coefficients, with values greater than 0.9 for 30 parameters and greater than 0.8 for seven parameters. The left heel bone angle expressed a moderate correlation, with a value of 0.609. The SEM values varied between 0.31 and 3.51 mm for the length and width measures, between 0.74 and 5.58 mm for the height data, between 0.75 and 5.9 mm for the circumferences, and between 0.78° and 2.98° for the angles. The smallest detectable difference values ranged from 0.86 to 16.36 mm for length, width, height, and circumference measures and from 2.17° to 8.26° for the angle measures. For the validity of the INFOOT three-dimensional foot digitizer, Pearson correlation coefficients varied between 0.750 and 0.997.
CONCLUSIONS: In this rheumatoid arthritis population, good validity was demonstrated compared with clinical measurements, and most of the obtained parameters proved to be reliable.

Entities:  

Mesh:

Year:  2011        PMID: 21622631     DOI: 10.7547/1010198

Source DB:  PubMed          Journal:  J Am Podiatr Med Assoc        ISSN: 1930-8264


  4 in total

1.  Approach for measuring the angle of hallux valgus.

Authors:  Jin Zhou; Petr Hlavacek; Bo Xu; Wuyong Chen
Journal:  Indian J Orthop       Date:  2013-05       Impact factor: 1.251

2.  Comparing 3D foot scanning with conventional measurement methods.

Authors:  Yu-Chi Lee; Gloria Lin; Mao-Jiun J Wang
Journal:  J Foot Ankle Res       Date:  2014-10-25       Impact factor: 2.303

3.  Reproducibility of foot dimensions measured from 3-dimensional foot scans in children and adolescents with Down syndrome.

Authors:  Nirmeen M Hassan; Andrew K Buldt; Nora Shields; Karl B Landorf; Hylton B Menz; Shannon E Munteanu
Journal:  J Foot Ankle Res       Date:  2020-06-04       Impact factor: 2.303

4.  Reliability and validity of the Microsoft Kinect for evaluating static foot posture.

Authors:  Benjamin F Mentiplay; Ross A Clark; Alexandra Mullins; Adam L Bryant; Simon Bartold; Kade Paterson
Journal:  J Foot Ankle Res       Date:  2013-04-08       Impact factor: 2.303

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.