Improved partition efficiency with threaded cylindrical column in vortex counter-current chromatography.

Yoichiro Ito1, Zhiyong Ma, Robert Clary, Jimmie Powell, Martha Knight, Thomas M Finn.   

Abstract

Type-I coil planet centrifuge produces a uniformly circulating centrifugal force field to produce vortex motion of two immiscible solvent phases in a cylindrical cavity of the separation column to perform efficient countercurrent chromatography. The partition efficiency obtained from the original vortex column was substantially improved by threading the cylindrical cavity to increase the area of mass transfer between the two phases. Partition efficiency of the threaded column was evaluated by three different two-phase solvent systems with a broad range of hydrophobicity each with a set of suitable test samples. Overall results of the present studies indicated that the threaded cylindrical column substantially improves the partition efficiency in terms of theoretical plate number, peak resolution, and height equivalent of one theoretical plate. The results also indicated that higher peak resolution is produced by eluting either the upper phase in the head to tail direction or the lower phase in the reversed direction. When there is a choice in the mobile phase, a better separation is achieved by using the less viscous phase as the mobile phase. Since the present system gives extremely low column pressure, it may be a potential alternative to the conventional type-J HSCCC system for a large-scale preparative separation. Published by Elsevier B.V.

Entities:  

Mesh:

Substances:

Year:  2011        PMID: 21616497      PMCID: PMC3109182          DOI: 10.1016/j.chroma.2011.04.081

Source DB:  PubMed          Journal:  J Chromatogr A        ISSN: 0021-9673            Impact factor:   4.759


  3 in total

1.  Band broadening inside the chromatographic column: the interest of a liquid stationary phase.

Authors:  Alain Berthod
Journal:  J Chromatogr A       Date:  2006-06-06       Impact factor: 4.759

2.  Evaluation of the effect of column orientation in type-I high-speed counter-current chromatography.

Authors:  Yi Yang; Dongyu Gu; Haji Akber Aisa; Yoichiro Ito
Journal:  J Chromatogr A       Date:  2010-03-09       Impact factor: 4.759

3.  Countercurrent chromatography with flow-through coil planet centrifuge.

Authors:  Y Ito; R L Bowman
Journal:  Science       Date:  1971-07-30       Impact factor: 47.728

  3 in total
  5 in total

1.  Spiral Tube Assembly for High-Speed Countercurrent Chromatography: Choice of Elution Modes for Four Typical Two-Phase Solvent Systems.

Authors:  Y Ito; R Clary; J Powell; M Knight; T M Finn
Journal:  J Liq Chromatogr Relat Technol       Date:  2008-01-01       Impact factor: 1.312

Review 2.  Countercurrent Separation of Natural Products: An Update.

Authors:  J Brent Friesen; James B McAlpine; Shao-Nong Chen; Guido F Pauli
Journal:  J Nat Prod       Date:  2015-07-15       Impact factor: 4.050

3.  Preparation of two novel monobrominated 2-(2',4'-dihydroxybenzoyl)-3,4,5,6-tetrachlorobenzoic acids and their separation from crude synthetic mixtures using vortex counter-current chromatography.

Authors:  Adrian Weisz; Jacob J Witten; Yun Zeng; Eugene P Mazzola; Yoichiro Ito
Journal:  J Chromatogr A       Date:  2012-03-19       Impact factor: 4.759

4.  Vortex Counter-Current Chromatography: Performance of a New Preparative Column.

Authors:  Yoichiro Ito; Robert Clary; Jacob J Witten; Yun Zeng
Journal:  Chromatographia       Date:  2012-10-01       Impact factor: 2.044

5.  Bioactivity-Guided Separation of Potential D₂ Dopamine Receptor Antagonists from Aurantii Fructus based on Molecular Docking Combined with High-Speed Counter-Current Chromatography.

Authors:  Yingjie He; Shihao Zhu; Changqiao Wu; Ying Lu; Qi Tang
Journal:  Molecules       Date:  2018-11-29       Impact factor: 4.411

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.