Biao Jin1, Christine Laskov, Massimo Rolle, Stefan B Haderlein.
Abstract
Compound-specific online chlorine isotope analysis of chlorinated hydrocarbons was evaluated and validated using gas chromatography coupled to a regular quadrupole mass spectrometer (GC-qMS). This technique avoids tedious off-line sample pretreatments, but requires mathematical data analysis to derive chlorine isotope ratios from mass spectra. We compared existing evaluation schemes to calculate chlorine isotope ratios with those that we modified or newly proposed. We also tested systematically important experimental procedures such as external vs. internal referencing schemes, and instrumental settings including split ratio, ionization energy, and dwell times. To this end, headspace samples of tetrachloroethene (PCE), trichloroethene (TCE), and cis-dichloroethene (cDCE) at aqueous concentrations in the range of 20-500 μg/L (amount on-column range: 3.2-115 pmol) were analyzed using GC-qMS. The results (³⁷Cl/³⁵Cl ratios) showed satisfying to good precisions with relative standard deviations (n = 5) between 0.4‰ and 2.1‰. However, we found that the achievable precision considerably varies depending on the applied data evaluation scheme, the instrumental settings, and the analyte. A systematic evaluation of these factors allowed us to optimize the GC-qMS technique to determine chlorine isotope ratios of chlorinated organic contaminants.Entities:
Mesh:
Substances:
Year: 2011 PMID: 21612209 DOI: 10.1021/es200749d
Source DB: PubMed Journal: Environ Sci Technol ISSN: 0013-936X Impact factor: 9.028