Literature DB >> 21610540

Computer-assisted surgery and intraoperative three-dimensional imaging for screw placement in different pelvic regions.

Lars Grossterlinden1, Jakob Nuechtern, Philipp G C Begemann, Ina Fuhrhop, Jan P Petersen, Andreas Ruecker, Martin Rupprecht, Wolfgang Lehmann, Udo Schumacher, Johannes M Rueger, Daniel Briem.   

Abstract

BACKGROUND: The purpose of this study was to assess the feasibility and accuracy of computer-assisted surgery (CAS) for screw placement in different pelvic regions using intraoperative three-dimensional (3D) imaging and to evaluate the influence of surgeons' experience with such a system on procedure time, radiation time, radiation dose, and misplacement rate.
METHODS: Experimental study in a human cadaveric model (n=5) for percutaneous screw placement in the anterior column of the acetabulum, the posterior pelvic ring (S1, S2), and the superior pubic ramus via 3D fluoroscopic navigated procedure. Accuracy of screw placement was assessed by 3D image intensifier, including the reconstruction of multiplanar images and by computer tomography (CT) scan. Influence of surgeons' experience was assessed by direct comparison of a low- and high-volume surgeon using the same technical setting.
RESULTS: In 100% of all procedures, intraoperative Iso-C3D image analysis was sufficient to confirm a correct screw placement. The postoperative CT scan revealed no further screw misplacement. However, for a correct supraacetabular screw placement, the intraoperative 3D scan was essential. In this group, the 3D scan showed screw misplacement in three cases. Procedure time for all indications and screw failure rate were significantly lower for the higher experienced surgeon.
CONCLUSION: The 3D fluoroscopic navigated procedure in pelvic surgery seems to be a useful tool for all surgeons and especially for less experienced ones. Furthermore, the intraoperative reconstruction of multiplanar 3D images allows a secure control of implant positioning.

Entities:  

Mesh:

Year:  2011        PMID: 21610540     DOI: 10.1097/TA.0b013e31820333dd

Source DB:  PubMed          Journal:  J Trauma        ISSN: 0022-5282


  3 in total

1.  Three-dimensional navigation-guided percutaneous screw fixation for nondisplaced and displaced pelvi-acetabular fractures in a major trauma centre.

Authors:  King Him Chui; Chi Chiu Dennis Chan; Ka Chun Ip; Kin Bong Lee; Wilson Li
Journal:  Int Orthop       Date:  2017-10-23       Impact factor: 3.075

2.  [Arthroscopically controlled screw placement for osteosynthesis of acetabular fractures].

Authors:  L-P Götz; R Schulz
Journal:  Unfallchirurg       Date:  2013-11       Impact factor: 1.000

3.  Intraoperative 3D imaging in the treatment of elbow fractures--a retrospective analysis of indications, intraoperative revision rates, and implications in 36 cases.

Authors:  Marc Schnetzke; Julia Fuchs; Sven Y Vetter; Nils Beisemann; Holger Keil; Paul-Alfred Grützner; Jochen Franke
Journal:  BMC Med Imaging       Date:  2016-03-18       Impact factor: 1.930

  3 in total