| Literature DB >> 21607889 |
Sunwoo Park1, Kihong Ryu, Jungyoon Kim, Jongsang Son, Youngho Kim.
Abstract
In this study, we have analysed heel strike (HS) and toe off (TO) of normal individuals and hemiplegic patients, taking advantage of output curves acquired from various sensors, and verified the validity of sensor detection methods and their effectiveness when they were used for hemiplegic gaits. Gait phase detections using three different motion sensors were valid, since they all had reliabilities more than 95%, when compared with foot velocity algorithm. Results showed that the tilt sensor and the gyrosensor could detect gait phase more accurately in normal individuals. Vertical acceleration could detect HS most accurately in hemiplegic patient group A. The gyrosensor could detect HS and TO most accurately in hemiplegic patient groups A and B. The detection of TO from all sensor signals was valid in both the patient groups A and B. However, the vertical acceleration detected HS validly in patient group A and the gyrosensor detected HS validly in patient group B.Entities:
Mesh:
Year: 2011 PMID: 21607889 DOI: 10.1080/10255842.2011.575376
Source DB: PubMed Journal: Comput Methods Biomech Biomed Engin ISSN: 1025-5842 Impact factor: 1.763