A bench study of the effects of leak on ventilator performance during noninvasive ventilation.

Yoshitoyo Ueno1, Nobuto Nakanishi, Jun Oto, Hideaki Imanaka, Masaji Nishimura.   

Abstract

BACKGROUND: During noninvasive ventilation (NIV), leak around the mask may cause inadequate ventilatory support or patient-ventilator asynchrony such as auto-triggering. Some NIV ventilators may be better than others at compensating for leak.
METHODS: We bench studied 3 NIV ventilators (Vision, Carina, and Trilogy100) and 2 ICU ventilators (Puritan Bennett 840 and Evita XL) to assess how they coped with 2 leak levels and zero leak during NIV. With a 2-bellows-in-a-box lung model we simulated spontaneous breathing with tidal volumes of 300 mL and 500 mL, at pressure support of 0 and 10 cm H(2)O and PEEP of 5 and 10 cm H(2)O. We affixed the airway opening of the lung model to the mouth of a mannequin head and secured a mask on the mannequin face. We created a medium leak and a large leak with different size holes, and measured PEEP in the presence of leak. We also measured the actual pressure-support values and calculated the deviations from the set pressure-support value and the pressure-time product (PTP) of the airway opening pressure below and above baseline.
RESULTS: With the medium leak only the Vision and Carina maintained the set PEEP and pressure support. With the large leak the pressure support was decreased with all the tested ventilators. With the larger leak and pressure support of 10 cm H(2)O the PTP below baseline for triggering increased with 2 ventilators, and the PTP above baseline for supporting the patient's inspiratory effort decreased with all 5 ventilators. The larger tidal volume increased the PTP below baseline with all 5 ventilators and at all leak sizes.
CONCLUSIONS: Some of the ventilators compensated for leak better than others. With the larger leak none of the ventilators maintained the set PEEP or pressure support.

Entities:  

Mesh:

Year:  2011        PMID: 21605483     DOI: 10.4187/respcare.01145

Source DB:  PubMed          Journal:  Respir Care        ISSN: 0020-1324            Impact factor:   2.258


  4 in total

1.  Bench-test comparison of 26 emergency and transport ventilators.

Authors:  Erwan L'Her; Annie Roy; Nicolas Marjanovic
Journal:  Crit Care       Date:  2014-10-15       Impact factor: 9.097

2.  Home noninvasive positive pressure ventilation with built-in software in stable hypercapnic COPD: a short-term prospective, multicenter, randomized, controlled trial.

Authors:  Luqian Zhou; Xiaoying Li; Lili Guan; Jianhua Chen; Bingpeng Guo; Weiliang Wu; Yating Huo; Ziqing Zhou; Zhenyu Liang; Yuqi Zhou; Jie Tan; Xin Chen; Yuanlin Song; Rongchang Chen
Journal:  Int J Chron Obstruct Pulmon Dis       Date:  2017-04-27

3.  Effectiveness of Inspiratory Termination Synchrony with Automatic Cycling During Noninvasive Pressure Support Ventilation.

Authors:  Yuqing Chen; Kewen Cheng; Xin Zhou
Journal:  Med Sci Monit       Date:  2016-05-20

4.  Partially RepRapable automated open source bag valve mask-based ventilator.

Authors:  Aliaksei Petsiuk; Nagendra G Tanikella; Samantha Dertinger; Adam Pringle; Shane Oberloier; Joshua M Pearce
Journal:  HardwareX       Date:  2020-08-11