Endotracheal tube extubation force: adhesive tape versus endotracheal tube holder.

Takeru Shimizu1, Taro Mizutani, Soichiro Yamashita, Keiichi Hagiya, Makoto Tanaka.   

Abstract

BACKGROUND: Adhesive tape is commonly used to secure the endotracheal tube (ETT) in anesthesia and intensive-care settings.
OBJECTIVE: To determine the force required to extubate when the ETT is secured with adhesive tape or commercially available ETT holders.
METHODS: We orally intubated a simulation manikin with a standard 8.0-mm inner-diameter ETT, inflated the cuff to 20 cm H(2)O, and measured the force required to extubate with the ETT secured in several ways. We tested 3 brands of tape (Durapore, Multipore Dry, and Wardel) with 6 methods, and 2 commercially available ETT holders (LockTite and Thomas) with one method. We also tested a bite block (Universal Bite Block) with 2 methods. We used a releasable cable tie with the bite block and/or ETT holder. We connected the ETT to a digital force gauge and pulled perpendicular to the oral cavity, until the entire cuff was removed from the trachea. In each trial we considered the largest force recorded the extubation force.
RESULTS: One of the conventional tape methods (with wider tape and longer tape strips) required the largest force to extubate.
CONCLUSIONS: With tape strips of sufficient length and width, a conventional tape method was superior to the 2 tested commercial ETT holders in holding the ETT in place in the manikin.

Mesh:

Year:  2011        PMID: 21605477     DOI: 10.4187/respcare.00954

Source DB:  PubMed          Journal:  Respir Care        ISSN: 0020-1324            Impact factor:   2.258


  3 in total

1.  A Comparison of the Haider Tube-Guard® Endotracheal Tube Holder Versus Adhesive Tape to Determine if This Novel Device Can Reduce Endotracheal Tube Movement and Prevent Unplanned Extubation.

Authors:  Jack C Buckley; Adam P Brown; John S Shin; Kirsten M Rogers; Nir N Hoftman
Journal:  Anesth Analg       Date:  2016-05       Impact factor: 5.108

2.  The effect of adhesive tape versus endotracheal tube fastener in critically ill adults: the endotracheal tube securement (ETTS) randomized controlled trial.

Authors:  Janna S Landsperger; Jesse M Byram; Bradley D Lloyd; Todd W Rice
Journal:  Crit Care       Date:  2019-05-07       Impact factor: 9.097

3.  Extubation force depends upon angle of force application and fixation technique: a study of 7 methods.

Authors:  Jennifer L Wagner; Robin Shandas; Craig J Lanning
Journal:  BMC Anesthesiol       Date:  2014-08-24       Impact factor: 2.217

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.