Biomechanical effect of bone cement augmentation on rotational stability and pull-out strength of the Proximal Femur Nail Antirotation™.

S Erhart1, W Schmoelz, M Blauth, A Lenich.   

Abstract

INTRODUCTION: After surgical treatment of osteoporotic hip fractures, complications such as implant cut-out are reported to be high and implant failure often is associated with poor bone quality. As augmentation is reported to enhance implant anchorage, the aim of our study was to investigate the effect of bone cement augmentation on the rotational stability and the pull-out resistance of the Proximal Femur Nail Antirotation™ (PFNa) blade.
MATERIALS AND METHODS: A total of 18 fresh-frozen femoral heads (mean age 68 years, standard deviation (SD) 8.2) were scanned with quantitative computed tomography (qCT) for bone mineral density (BMD) measurements and instrumented with a PFNa blade. Nine specimens were augmented with a mean volume of 4.4 ml Traumacem V+. After cement consolidation, the blade was rotated for 60° for the rotational test. Subsequently, the blade was extracted from the specimens. Force, torque, displacement and angle were recorded constantly.
RESULTS: In the rotational test, the mean maximum torque in the augmented group (17.2 Nm, SD 5.0) was significantly higher (p=0.017) than in the non-augmented group (11.7 Nm, SD 3.5). The pull-out test also yielded a significant difference (p=0.047) between the augmented (maximum pullout force: 2315.2N, SD 1060.6) and the non-augmented group (1180.4N, SD 1171.4). DISCUSSION: Augmentation of femoral heads yielded a significantly superior rotational stability, as well as an enhanced pull-out resistance, compared to the non-augmented state. However, the higher the BMD of the specimens, the lower was the effect of augmentation on the rotational stability. Therefore, augmentation can be a good clinical tool to enhance implant anchorage in osteoporotic bone.
Copyright © 2011 Elsevier Ltd. All rights reserved.

Entities:  

Mesh:

Substances:

Year:  2011        PMID: 21601203     DOI: 10.1016/j.injury.2011.04.010

Source DB:  PubMed          Journal:  Injury        ISSN: 0020-1383            Impact factor:   2.586


  15 in total

1.  Cement augmentation of the proximal femur nail antirotation: is it safe?

Authors:  Konrad Schuetze; S Ehinger; A Eickhoff; C Dehner; F Gebhard; P H Richter
Journal:  Arch Orthop Trauma Surg       Date:  2020-07-25       Impact factor: 3.067

Review 2.  [Pertrochanteric femoral fractures in the elderly].

Authors:  G H Sandmann; P Biberthaler
Journal:  Unfallchirurg       Date:  2015-05       Impact factor: 1.000

Review 3.  Intertrochanteric fractures: a review of fixation methods.

Authors:  Senthil Nathan Sambandam; Jayadev Chandrasekharan; Varatharaj Mounasamy; Cyril Mauffrey
Journal:  Eur J Orthop Surg Traumatol       Date:  2016-03-30

Review 4.  [Osteoporotic fractures of the proximal femur. What's new?].

Authors:  F Bonnaire; C Straßberger; M Kieb; P Bula
Journal:  Chirurg       Date:  2012-10       Impact factor: 0.955

Review 5.  Cement augmentation for trochanteric fracture in elderly: A systematic review.

Authors:  Leonardo Stramazzo; Salvatore Ratano; Francesco Monachino; Davide Pavan; Giuseppe Rovere; Lawrence Camarda
Journal:  J Clin Orthop Trauma       Date:  2020-10-20

6.  Indications for cement augmentation in fixation of geriatric intertrochanteric femur fractures: a systematic review of evidence.

Authors:  L Henry Goodnough; Harsh Wadhwa; Seth S Tigchelaar; Malcolm R DeBaun; Michael J Chen; Matt L Graves; Michael J Gardner
Journal:  Arch Orthop Trauma Surg       Date:  2021-04-07       Impact factor: 2.928

7.  The PFNA® Augmented in Revision Surgery of Proximal Femur Fractures.

Authors:  Alexander Scola; Florian Gebhard; Christoph Dehner; Götz Röderer
Journal:  Open Orthop J       Date:  2014-07-11

8.  Osteosynthesis-screw augmentation by ultrasound-activated biopolymer--an ovine in vivo study assessing biocompatibility and bone-to-implant contact.

Authors:  Hanjo Neumann; Stefan Breer; Nils Reimers; Richard Kasch; Arndt-Peter Schulz; Benjamin Kienast
Journal:  J Orthop Surg Res       Date:  2015-01-28       Impact factor: 2.359

9.  Early Experience with the Trochanteric Fixation Nail-Advanced (TFN-A): A Descriptive Review of Thirty-Four Cases from a Single Center.

Authors:  Joshua De Castro Unsay; Ivan Chua Tjun Huat; Ernest Kwek Beng Kee
Journal:  Indian J Orthop       Date:  2020-08-20       Impact factor: 1.251

Review 10.  Management of proximal femur fractures in the elderly: current concepts and treatment options.

Authors:  H Fischer; T Maleitzke; C Eder; S Ahmad; U Stöckle; K F Braun
Journal:  Eur J Med Res       Date:  2021-08-04       Impact factor: 2.175

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.