Literature DB >> 21600892

A rapid and sensitive chemiluminescence immunoassay based on magnetic particles for squamous cell carcinoma antigen in human serum.

Huisheng Zhang1, Suwen Qi.   

Abstract

BACKGROUND: Because squamous cell carcinoma antigen (SCCa) quantification has demonstrated strong clinical potential, we describe a rapid and highly sensitive magnetic particle-based chemiluminescence immunoassay (CLIA) technique for assaying SCCa in serum.
METHODS: Fluorescein isothiocyanate (FITC) and N-(aminobutyl)-N-(ethylisoluminol) (ABEI) were used to label 2 different monoclonal antibodies of anti-SCCa. Both of the labeled antibodies combined with SCCa to form a sandwiched immunoreaction that was monitored by chemiluminescence (CL) detection. The magnetic particles (MPs) that were coated with anti-FITC antibody served as both the solid phase and the separator. The relevant variables involved in the CLIA signals were optimized and the parameters of the proposed method were evaluated.
RESULTS: The method was linear to 20 ng/ml SCCa with a detection limit of 0.02 ng/ml. The intra-assay imprecision results [mean (CV)] were 1.12 ng/ml (3.81%), 2.58 ng/ml (2.53%) and [6.56 ng/ml (2.24%)]; the inter-assay imprecision results were [1.18 ng/ml (5.26%)], [2.49 ng/ml (4.75%)] and [6.61 ng/ml (4.29%)]. The average recoveries were between 97% and 104%. The relationship between the concentration of diluted SCCa and the dilution ratios gave a linear correlation coefficient of 0.9995. A correlation analysis against an established automated assay generated a slope of 0.9929 and an intercept of 0.0039 ng/ml (r=0.9964).
CONCLUSIONS: The proposed method demonstrates an acceptable performance for quantifying serum SCCa and is suitable for the fabrication of a commercial kit with application in the automated CL analyzer.
Copyright © 2011 Elsevier B.V. All rights reserved.

Entities:  

Mesh:

Substances:

Year:  2011        PMID: 21600892     DOI: 10.1016/j.cca.2011.05.005

Source DB:  PubMed          Journal:  Clin Chim Acta        ISSN: 0009-8981            Impact factor:   3.786


  4 in total

1.  A simple and rapid capillary chemiluminescence immunoassay for quantitatively detecting human serum HBsAg.

Authors:  A Xiang; F Wei; X Lei; Y Liu; Y Liu; Y Guo
Journal:  Eur J Clin Microbiol Infect Dis       Date:  2013-06-19       Impact factor: 3.267

2.  Comparison of 1-Ethyl-3-(3-Dimethylaminopropyl) Carbodiimide Based Strategies to Crosslink Antibodies on Amine-Functionalized Platforms for Immunodiagnostic Applications.

Authors:  Sandeep Kumar Vashist
Journal:  Diagnostics (Basel)       Date:  2012-08-27

3.  A reusable localized surface plasmon resonance biosensor for quantitative detection of serum squamous cell carcinoma antigen in cervical cancer patients based on silver nanoparticles array.

Authors:  Qianying Zhao; Ruiqi Duan; Jialing Yuan; Yi Quan; Huan Yang; Mingrong Xi
Journal:  Int J Nanomedicine       Date:  2014-02-22

4.  A Novel Controlled Release Immunosensor based on Benzimidazole Functionalized SiO2 and Cyclodextrin Functionalized Gold.

Authors:  Hongmin Ma; Yaoguang Wang; Dan Wu; Yong Zhang; Jian Gao; Xiang Ren; Bin Du; Qin Wei
Journal:  Sci Rep       Date:  2016-01-21       Impact factor: 4.379

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.