Evaluation of the corneal endothelium using noncontact and contact specular microscopy.

Eszter Szalai1, Gábor Németh, András Berta, László Módis.   

Abstract

PURPOSE: To determine the difference between the endothelial cell density (ECD) and the morphologic measurements assessed with a recently developed noncontact specular microscope compared with a contact endothelial microscope. The agreement between the 2 independent operators was also investigated.
METHODS: The central corneal endothelium was examined in the right eyes of 41 healthy individuals using automated image analysis programs without correction for a noncontact specular microscope and a contact specular microscope (EM-3000, EM-1000; Tomey, Tennenlohe, Germany). The evaluated parameters such as ECD,average cell area (AVG), and coefficient of variation of the cell area were determined by 2 operators. Conversion factors were also considered to adjust the ECD values.
RESULTS: ECD measurements obtained with the noncontact instrument (2734 ± 287 cells per square millimeter, first operator; 2726 ± 282 cells per square millimeter, second operator) were significantly higher (P < 0.0001) in comparison with the contact microscope (2283 ± 206 cells per square millimeter, first operator; 2274 ± 242 cells per square millimeter, second operator) by both investigators. The contact specular microscope significantly overestimated (P < 0.0001) AVG and coefficient of variation of the cell area values. Statistically significant correlation (P < 0.0001) was detected in ECD (r = 0.77, first operator) and AVG measurements (r = 0.72, first operator) between the 2 endothelial microscopes. Higher degree of agreement between operators was given when using the noncontact specular microscope compared with the contact instrument.
CONCLUSIONS: The significant differences and the poor agreement between the automated software options disclosed that the 2 instruments cannot be used interchangeably.

Mesh:

Year:  2011        PMID: 21598434     DOI: 10.1097/ico.0b013e3182000807

Source DB:  PubMed          Journal:  Cornea        ISSN: 0277-3740            Impact factor:   2.651


  7 in total

1.  Assessment of a variable frame (polygonal) method to estimate corneal endothelial cell counts after corneal transplantation.

Authors:  S Jonuscheit; M J Doughty; K Ramaesh
Journal:  Eye (Lond)       Date:  2012-03-23       Impact factor: 3.775

2.  Measurements of central corneal thickness and endothelial parameters with three different non-contact specular microscopy devices.

Authors:  Ozgur Cakici; Remzi Karadag; Huseyin Bayramlar; Efe Koyun
Journal:  Int Ophthalmol       Date:  2016-05-24       Impact factor: 2.031

3.  Comparison of semi-automated center-dot and fully automated endothelial cell analyses from specular microscopy images.

Authors:  Sachiko Maruoka; Shunsuke Nakakura; Naoko Matsuo; Kayo Yoshitomi; Chikako Katakami; Hitoshi Tabuchi; Taiichiro Chikama; Yoshiaki Kiuchi
Journal:  Int Ophthalmol       Date:  2017-10-30       Impact factor: 2.031

4.  Low-Cost, Smartphone-Based Specular Imaging and Automated Analysis of the Corneal Endothelium.

Authors:  Sreekar Mantena; Jay Chandra; Eryk Pecyna; Andrew Zhang; Dominic Garrity; Stephan Ong Tone; Srinivas Sastry; Madhu Uddaraju; Ula V Jurkunas
Journal:  Transl Vis Sci Technol       Date:  2021-04-01       Impact factor: 3.283

5.  Qualitative Alterations on Corneal Endothelial Cell Morphometry and Hexagonality After Cataract Surgery.

Authors:  Francesco Saverio Sorrentino
Journal:  Clin Ophthalmol       Date:  2021-12-29

6.  A Pilot Study to Propose a "Harm Scale", a New Method to Predict Risk of Harm to the Corneal Endothelium Caused by Longitudinal Phacoemulsification, and the Subsequent Effect of Endothelial Damage on Post Operative Visual Acuity.

Authors:  Francesco Saverio Sorrentino; Claudio Bonifazzi; Francesco Parmeggiani; Paolo Perri
Journal:  PLoS One       Date:  2016-01-13       Impact factor: 3.240

7.  Torsional phacoemulsification: A pilot study to revise the "harm scale" evaluating the endothelial damage and the visual acuity after cataract surgery.

Authors:  Francesco Saverio Sorrentino; Silvia Matteini; Aurelio Imburgia; Claudio Bonifazzi; Adolfo Sebastiani; Francesco Parmeggiani
Journal:  PLoS One       Date:  2017-10-26       Impact factor: 3.240

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.