Literature DB >> 21576036

Comparison of two servo ventilator devices in the treatment of complex sleep apnea.

Tomasz J Kuzniar1, Smita Patel, Cynthia L Nierodzik, Loretta C Smith.   

Abstract

OBJECTIVE: Servo ventilation (SV) devices generate positive airway pressure with a variable pressure support that changes in response to a patient's own respiratory output. Two currently available SV devices-VPAP-AdaptSV® and BIPAP-AutoSV®-have been used in treatment of complex sleep apnea (CompSAS), but no side-by-side comparisons are available.
METHODS: Data of 76 consecutive patients with complex sleep apnea, who were prescribed a VPAP-AdaptSV® or BIPAP-AutoSV® in a non-randomized parallel design, were retrospectively analyzed. Patients underwent a diagnostic polysomnogram followed by a continuous positive airway pressure (CPAP) titration and a SV titration study. Objective compliance with the device was assessed at the first visit at 4-6weeks of its use.
RESULTS: Thirty-five patients received a VPAP-AdaptSV® device, while 41 patients were treated with BIPAP-AutoSV®. Patients treated with BIPAP-AutoSV® had a significantly higher apnea-hypopnea index during their CPAP titration study than patients treated with VPAP-AdaptSV® [49/h (28-60) vs. 35/h (19.5-49.5), median (interquartile range), p<0.001]. On follow-up, 56 patients (73.7%) were using their device. Mean nightly use was 5.0h (2.8-6.4) for VPAP-AdaptSV® group and 6.0h (3.5-7.2) for BIPAP-AutoSV® group (p=0.081); an improvement in Epworth Sleepiness Scale score was higher in the BIPAP-AutoSV® group than in the VPAP-AdaptSV® group [4 (1-9) vs. 2.5 (0-5), p=0.02].
CONCLUSION: Our retrospective data indicate that the two servo-ventilation devices are comparable means of controlling complex sleep apnea, and the compliance with them is high.
Copyright © 2011 Elsevier B.V. All rights reserved.

Entities:  

Mesh:

Year:  2011        PMID: 21576036     DOI: 10.1016/j.sleep.2010.09.017

Source DB:  PubMed          Journal:  Sleep Med        ISSN: 1389-9457            Impact factor:   3.492


  6 in total

1.  Complex sleep apnea unmasked by the use of a mandibular advancement device.

Authors:  Tomasz J Kuźniar; Ružica Kovačević-Ristanović; Thomas Freedom
Journal:  Sleep Breath       Date:  2010-12-29       Impact factor: 2.816

2.  The complex sleep apnea resolution study: a prospective randomized controlled trial of continuous positive airway pressure versus adaptive servoventilation therapy.

Authors:  Timothy I Morgenthaler; Tomasz J Kuzniar; Lisa F Wolfe; Leslee Willes; William C McLain; Rochelle Goldberg
Journal:  Sleep       Date:  2014-05-01       Impact factor: 5.849

3.  Analysis of cardiopulmonary coupling to assess adaptive servo-ventilation success in complex sleep apnea management.

Authors:  Kannan Ramar; Benoît Desrues; Priya Ramar; Timothy I Morgenthaler
Journal:  Sleep Breath       Date:  2012-11-02       Impact factor: 2.816

4.  Clinical heterogeneity of patients with complex sleep apnea syndrome.

Authors:  Tomasz J Kuźniar; Kamilla Kasibowska-Kuźniar; Daniel W Ray; Thomas Freedom
Journal:  Sleep Breath       Date:  2013-02-23       Impact factor: 2.816

5.  Complex sleep apnea syndrome.

Authors:  Juan Wang; Yan Wang; Jing Feng; Bao-Yuan Chen; Jie Cao
Journal:  Patient Prefer Adherence       Date:  2013-07-03       Impact factor: 2.711

Review 6.  Complex sleep apnea syndrome.

Authors:  Muhammad Talha Khan; Rose Amy Franco
Journal:  Sleep Disord       Date:  2014-02-16
  6 in total