Experience with the Fassier-Duval telescopic rod: first 24 consecutive cases with a minimum of 1-year follow-up.

Oliver Birke1, Neville Davies, Mark Latimer, David Graham Little, Michael Bellemore.   

Abstract

BACKGROUND: The new Fassier-Duval Telescopic IM System (FD-rod) has the advantage of a single entry point over the traditional telescopic rods such as the Bailey-Dubow or Sheffield rods. Although encouraging early results were presented by the originators of the technique at international meetings, there is no formal publication in the literature as yet.
METHODS: We performed a chart and x-ray review of the first 24 consecutive FD-rod insertions in 15 patients (age, 1.5 to 12.5 y) with a minimum of 1-year follow up (1 to 2.4 y) after implantation of femoral and/or tibial FD-rods. Diagnoses included with osteogenesis imperfecta (OI) (15 cases, 9 patients), and other conditions such as congenital tibial pseudarthrosis (CPT) in neurofibromatosis type 1 (NF1) (2 cases), and epidermal naevus syndrome (1 case). In patients with hypophosphataemic rickets (6 cases, 2 patients) the FD-rods were combined with an Ilizarov frame.
RESULTS: We found the OI patient group associated with a 13% reoperation rate (2 of 15 cases) for proximal rod migration and a 40% complication rate (6 of 15 cases): rod migration and limited telescoping (5) and intraoperative joint intrusion (1). There were no infections. All the NF1 CPT (2) and epidermal naevus syndrome (1) cases required several reoperations for nonunion, loss of fixation, shortening (negative telescoping), migration, and/or joint intrusion-mainly due to the severe underlying pathology with insufficient longitudinal or torsional stability and diminished healing capacity. In hypophosphataemic rickets (combined with Ilizarov frame fixation) we found a 50% complication rate (3 of 6 cases) and a 17% reoperation rate (1 of 6): 2 FD-rods did not telescope and 1 case of peroneal neuropraxia required neurolysis.
CONCLUSIONS: In our experience the technique of using FD rods is demanding and associated with some intraoperative and postoperative pitfalls. We are happy to continue its use in OI patients when there is longitudinal stability and sufficient bone healing. However, in circumstances of insufficient stability and bone healing potential, further stabilization that can be achieved with an Ilizarov frame may be beneficial.

Entities:  

Mesh:

Year:  2011        PMID: 21572286     DOI: 10.1097/BPO.0b013e31821bfb50

Source DB:  PubMed          Journal:  J Pediatr Orthop        ISSN: 0271-6798            Impact factor:   2.324


  21 in total

1.  CORR Insights®: Dual Interlocking Telescopic Rod Provides Effective Tibial Stabilization in Children With Osteogenesis Imperfecta.

Authors:  Thoralf R Liebs
Journal:  Clin Orthop Relat Res       Date:  2018-11       Impact factor: 4.176

Review 2.  Osteogenesis imperfecta: diagnosis and treatment.

Authors:  A Biggin; C F Munns
Journal:  Curr Osteoporos Rep       Date:  2014-09       Impact factor: 5.096

3.  Dual Interlocking Telescopic Rod Provides Effective Tibial Stabilization in Children With Osteogenesis Imperfecta.

Authors:  Chang Ho Shin; Doo Jae Lee; Won Joon Yoo; In Ho Choi; Tae-Joon Cho
Journal:  Clin Orthop Relat Res       Date:  2018-11       Impact factor: 4.176

4.  The clinical features of osteogenesis imperfecta in Vietnam.

Authors:  Ho Duy Binh; Katre Maasalu; Vu Chi Dung; Can T Bich Ngoc; Ton That Hung; Tran V Nam; Le N Thanh Nhan; Ele Prans; Ene Reimann; Lidiia Zhytnik; Sulev Kõks; Aare Märtson
Journal:  Int Orthop       Date:  2016-11-02       Impact factor: 3.075

5.  Combined technique of titanium telescopic rods and external fixation in osteogenesis imperfecta patients: First 12 consecutive cases.

Authors:  Dmitry Popkov; Tamara Dolganova; Eduard Mingazov; Dmitry Dolganov; Andrey Kobyzev
Journal:  J Orthop       Date:  2020-06-20

6.  Retro-patellar approach in telescopic nailing of the tibia in children with osteogenesis imperfecta.

Authors:  Ozan A Erdal; Baris Gorgun; Ilker A Sarikaya; Muharrem Inan
Journal:  J Child Orthop       Date:  2021-08-20       Impact factor: 1.548

Review 7.  The Ilizarov paradigm: thirty years with the Ilizarov method, current concerns and future research.

Authors:  Alexander V Gubin; Dmitry Y Borzunov; Tatiana A Malkova
Journal:  Int Orthop       Date:  2013-05-28       Impact factor: 3.075

8.  [Osteogenesis imperfecta].

Authors:  T Wirth
Journal:  Orthopade       Date:  2012-09       Impact factor: 1.087

9.  Fassier-Duval Rod Failure: Is It Related to Positioning in the Distal Epiphysis?

Authors:  Kaley Holmes; Jane Gralla; Christopher Brazell; Patrick Carry; Suhong Tong; Nancy H Miller; Gaia Georgopoulos
Journal:  J Pediatr Orthop       Date:  2020-09       Impact factor: 2.537

10.  Early telescopic rod osteosynthesis for Osteogenesis Imperfecta patients.

Authors:  A Sterian; R Balanescu; A Barbilian; I Tevanov; M Carp; C Nahoi; M Barbu; A Ulici
Journal:  J Med Life       Date:  2015 Oct-Dec
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.